9 TROUBLESHOOTING
Service Indicator Light Diagnostics
For this procedure, refer to FIGURE 1.
|
| Service Indicator Light |
|
| FIGURE 1 Service Indicator Light Diagnostics |
NUMBER OF | ERROR CODE | POSSIBLE SOLUTION |
FLASHES | DESCRIPTION |
|
1 | User Fault | Release joystick to neutral and try again. |
2 | Battery Fault | Charge the batteries. Refer to Charging Batteries on page 53. |
|
| Check that battery cables are connected properly. If necessary, replace batteries. Refer to |
|
| Replacing the Batteries and/or Battery Cables on page 83. |
3 | Left Motor Fault | Contact Invacare/Dealer for service. |
4 | Right Motor Fault | Contact Invacare/Dealer for service. |
5 | Left Park Brake Fault | Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is |
|
| plugged into the controller. |
|
| Contact Invacare/Dealer for service. |
6Right Park Brake Fault Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is plugged into the controller.
Contact Invacare/Dealer for service.