When the control solution test result is within the control range, you are ready to test your blood. Your con- trol solution test will be stored in the meter memory. Record control solution tests in your logbook by placing a C next to the result to indicate a control solution test.

 

NOTES

TEST

 

DATE

 

TEST

 

RESULT

 

TIME

NOTES

INSULIN/

 

MEDICATION

 

TEST

 

DATE

 

TEST

 

RESULT

 

TIME

 

INSULIN/

 

MEDICATION

 

Control Solution Range

If test results fall outside the expected range, repeat the test. Results that fall outside the expected range may indicate the system is not working properly. Some possible causes are:

Procedural error

Expired or contaminated control solution

Test strip code and meter code do not match

Dirty meter or test strip holder

Test strip deterioration

Meter malfunction

Control solution that is outside of the temperature range 59–95˚F (15–35˚C)

If you are unable to resolve the problem, call LifeScan Customer Services, 1 800 227-8862.

CHECKING

17

Page 25
Image 25
Lifescan blood glucose monitoring system manual Control Solution Range