Confirmation Dot

Error Messages

The meter gives specific error mes- sages when it detects a problem with its internal functions, the test strip, or the testing procedure. However, improper use may cause an inaccurate

result without producing an error message or a symbol. When error mes- sages appear, note the error number, turn off the meter, then resolve the error by following these instructions:

ERROR

POSSIBLE CAUSE

WHAT TO DO

 

1. Blood was applied to

1. Clean the meter. Repeat the test

 

 

 

 

 

 

 

the wrong side of the

with a new test strip, applying

 

 

 

 

 

 

 

 

test strip, or

blood to the pink test square.

 

 

 

 

 

 

 

 

 

2. Not enough blood is

2. Repeat the test with a new test

 

on the test strip, or

strip. Apply a larger drop of blood

 

 

to the pink test square and

 

 

verify that the confirmation

 

 

dot on the back of the test strip

 

 

has turned completely blue.

Pink

Test

Square

MESSAGES

59