Mitsubishi Electronics FX1S manual Device specification, Setting range, Direction setting

Models: FX1S

1 120
Download 120 pages 31.35 Kb
Page 100
Image 100

FX Series Programmable Controllers

Devices in Detail 4

4.11.31 Phase Counters - User Start and Reset (C235 - C240)

These counters only use one input each. When direction flag M8235 is ON, counter C235 counts down. When it is OFF, C235 counts up.

When X11 is ON, C235 resets to 0 (zero). All contacts of the counter C235 are also reset. When X12 is ON, C235 is selected. From the previous counter tables, the corresponding counted input for C235 is X0. C235 therefore counts the number of times X0 switches from OFF to ON.

X10

X11

X12

M8235

RST C235

C235

K1234

Device specification:

• All of these counters are 32bit up/down ring counters. Their counting and contact operations are the same as normal 32bit up/down counters described on page 4-21. When the counters current value reaches its maximum or setting value, the counters associated contacts are set and held when the counter is counting upwards. However, when the counter is counting downwards the contacts are reset.

Setting range:

-2,147,483,648 to +2,147,483,647

Direction setting:

The counting direction for 1 phase counters is dependent on their corresponding flag M8✰✰✰; where ✰✰✰ is the number of the corresponding counter, (C235 to C240). When M8✰✰✰ is ON the counter counts down,

When M8✰✰✰ is OFF the counter counts up.

Using the SPD instruction:

Care should be taken when using the SPD applied instruction (FNC 56). This instruction has both high speed counter and interrupt characteristics, therefore input devices X0 through X5 may be used as the source device for the SPD instruction. In common with all high speed processes the selected source device of the SPD instruction must not coincide with any other high speed function which is operating, i.e. high speed counters or interrupts using the same input.

When the SPD instruction is used it is considered by the system to be a 1 phase high speed counter. This should be taken into account when summing the maximum com- bined input signal frequencies - see the previous section.

4-26

Page 100
Image 100
Mitsubishi Electronics FX1S manual Device specification, Setting range, Direction setting, Using the SPD instruction