Motorola i860 manual To connect the unit to the R-2660, Coaxial Cable In/Out Type

Models: i860

1 106
Download 106 pages 45.2 Kb
Page 64
Image 64

PREPARING FOR FIELD LEVEL TESTING: Connecting an iDEN Unit to the R-2660

Coaxial Cable

RF

In/Out

 

 

(N-type)

Unit Under Test

R-2660

Communications

System

BatteryAnalyzer

Eliminator

+

5.0to 12.0 Vdc Power Supply

Figure 7-1. Typical R-2660 Setup

To connect the unit to the R-2660:

1.Ensure the unit is powered off, and then turn on the R-2660.

2.Remove the battery cover and battery from the unit, and insert the reference SIM card.

3.Attach the battery eliminator to the back of the unit.

4.Engage RF.

5.Connect the SMA connector of the RF cable to the RF connector on the battery eliminator.

6.Connect the N-type connector of the RF cable to the R-2660 RF In/Out connector.

7.Attach the power leads on the battery eliminator to the DC power supply.

FCAUTION: Be very careful to observe polarity when connecting power to the battery eliminator.

8.Turn on the power supply, and adjust output between 5.0 and 12.0 Vdc. With the correct setting, the regulated battery eliminator will provide 4.0 Vdc to the subject unit.

9.If power supply is capable, set current limit = 4.0 A.

68P80400P73-A

7-3

Page 64
Image 64
Motorola i860 manual To connect the unit to the R-2660, Coaxial Cable In/Out Type