TARGET Series

ENMET Corporation

6.4 Advanced Maintenance Menu, Software 4.94 version

The Advanced Maintenance menu provides access to enhanced programming features of the TARGETinstrument.

To enter the Advanced Maintenance menu, press the MENU button repeatedly from the main gas display or Operation menu. Move the asterisk until it is next to the Maintenance menu line. Press the SELECT button to move to the password entry screen. The password is set to 1270 from the factory. Press and hold the MENU button for about two seconds. Note that the third line changes from STD Maintenance Menu to ADV Maintenance Menu. Use the MENU button to increment the first digit and the SELECT button to move the cursor to the next digit. Press SELECT when the proper password has been entered. If proper password is not entered, INVALID PASSWORD is displayed and the instrument returns to Operation menu.

Enter Password

to access

STD MAINTENANCE MENU

1270

*Calibrate Set Alarms Enable Sensors Set Clock/Chirp

Press and hold MENU

MENU

Enter Password

to access

ADV MAINTENANCE MENU

1270

*Set

Password

 

Set

Battery

 

Set

Op. Menu Type

Pump Cal

Exit

 

 

 

Example: Display of Advanced Maintenance Menu

6.4.1 Calibration

The calibration procedure in the Advanced Maintenance menu is very similar to the procedure in the Standard Maintenance menu, section 6.3. The difference is that in the Advanced Maintenance menu version, the span values can be modified and the zero and span settings must be acknowledged by pressing the SELECT button.

*Calibrate Set Alarms Enable Sensors Set Clock/Chirp

SELECT

ADV CALIBRATION MENU *BRH O2/CO/H2S/Mth Individual Sensors

Example: Display of Advanced Calibration Menu

If the BRH sensor is installed, see section 6.3.3 for applying span gas.

Expose the instrument to fresh air. When the mV reading on the bottom line is stable, press the SELECT button to set the ZERO reading.

CAL: BRH *ZERO: 0 ppm SPAN: 100 ppm BRH: 196 mV

SELECT

CAL: BRH

ZERO: 0 ppm *SPAN: 100 ppm BRH: 1504 mV

Example: Display of BRH Calibration

The asterisk now highlights the SPAN gas concentration. Default values for each sensor are listed in Section 8.0.

To modify the span value, use the Menu key to initiate the change and then increment the digits. The SELECT button is used to select the digit. SELECT the last digit only after span gas has been applied and the mV reading is stable.

If no modification is necessary, apply calibration gas and press the SELECT button when the mV reading is stable.

Repeat the procedure for each sensor as necessary.

6.4.2 Setting Instantaneous Alarms

From the Advanced Maintenance menu, press the MENU button once to put the asterisk next to the Set Alarms line. Press the SELECT button to enter the Set Alarms screen. Use the MENU button to move the asterisk next to the desired gas for alarm to be changed. Press the SELECT button to move the cursor to the first digit. The MENU button is used to increment the digit and the SELECT button to move the cursor to the next digit. Press the SELECT

15

Page 18
Image 18
O2 Innovations pmn manual Advanced Maintenance Menu, Software 4.94 version, Example Display of Advanced Maintenance Menu