TARGET Series | ENMET Corporation |
3.4.3 Over-range and Under-range Alarms
The
The
3.4.4 Low Battery
The TARGET Series instrument has a low battery alarm to provide the user with adequate warning of a low battery condition. The instrument provides a steady tone audio alarm when there is approximately 15 to 30 minutes of battery life left. A low battery warning message is displayed on the status line of the instrument LCD. The audio alarm can be acknowledged by pressing the SELECT button. The audio alarm and backlight reactivate after approximately five minutes. See section 3.2 for battery charging procedure.
3.4.5 Fault
Fault alarms have a steady audio alarm and are displayed on the status line of the LCD. The status line describes the fault condition and appropriate action should be take to resolve the issue. See section 11 for troubleshooting information.
3.5 Pop-Up Menu
The
Example: Pop-Up Menu
NOTE: The
3.5.1 Auto-Zero
To compensate for sensor aging and drift, the TARGETinstrument has an
environment.
To perform an
SELECT | O2 | : | OK |
| CO | : | OK |
| H2S: | OK | |
| Mth: | OK BRH: OK | |
|
|
|
|
Example: Display, Auto-Zero from Pop-Up Menu
7