TARGET Series

ENMET Corporation

9.4 Advanced Maintenance Menu, Software 6.0 version

*Calibrate

 

 

ADV CALIBRATION MENU

Set Alarms

 

 

*BRH

Enable Sensors

 

O2/CO/H2S/Mth

Pump Cal

Exit

 

Individual Sensors

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CALIBRATION MENU

 

 

 

 

*Exit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL: BRH

 

 

 

 

 

 

 

*ZERO:

0 ppm

 

 

 

 

 

 

 

 

SPAN:

100 ppm

 

 

 

 

 

 

 

 

BRH:

196 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL: BRH

 

 

 

To change

 

ZERO:

0 ppm

 

 

 

 

*SPAN:

100 ppm

 

 

 

zero gas

 

 

 

 

 

 

 

BRH:

1504 mV

 

 

 

value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To change

complete

 

 

 

 

 

 

Calibrate

*Set Alarms

Enable Sensors

Pump Cal Exit

SET ALARMS

*O2 19.5 / 23.5 % CO 35 ppm

H2S 10 ppm

SET ALARMS Mth50%LEL BRH 100 ppm *Exit

SET ALARMS

span gas

*O2

19.5 / 23.5 %

value

 

 

 

CO

35 ppm

 

H2S 10 ppm

 

Calibrate

Set Alarms

*Enable Sensors

Pump Cal Exit

Calibrate

Set Alarms

Enable Sensors

*Pump Cal Exit

Calibrate

Set Alarms

Enable Sensors

Pump Cal *Exit

ENABLE SENSORS *O2 ON

CO ON

H2S ON

ENABLE SENSORS

*Mth ON

BRH ON

Exit

*Pump Calibration

1st PT: Clear Flow

Menu: Exit

mV Sample

02 20.9%

CO 0ppm H2S 0ppm Mth0%LEL BRH 0ppm

ENABLE SENSORS *O2 OFF

CO ON

H2S ON

*Block Pump Inlet*

*Pump Calibration

2nd PT:Blocked Flow

Set:Fwrd Menu:Exit

mV Sample

Main Gas

Display

=SELECT

=MENU

29

Page 32
Image 32
O2 Innovations pmn manual Advanced Maintenance Menu, Software 6.0 version