O2 Innovations pmn Troubleshooting, TARGET Series, ENMET Corporation, Symptom, Possible Cause

Models: pmn

1 36
Download 36 pages 1.53 Kb
Page 33
Image 33
10.0 Troubleshooting

TARGET Series

ENMET Corporation

10.0 Troubleshooting

Symptom

Possible Cause

Action

Unit will not turn on.

Dead battery

Plug instrument into charger. Check

 

 

charge status.

 

 

 

Sensor fails calibration

Failed sensor

Replace sensor and recalibrate

 

Wrong gas concentration

Use proper calibration gas.

 

Old calibration gas

Replace calibration gas.

 

 

 

Calibration due on status line

Sensor hasn’t been calibrated in last 30

Check Cal Status in Pop-Up menu.

 

days.

Calibrate affected sensor(s).

 

 

 

Cal error on status line

Sensor failed a calibration.

Check Cal Status in Pop-Up menu.

 

 

Calibrate affected sensor(s). Replace

 

 

sensor if necessary and recalibrate.

 

 

 

Setup/config error on status line

Memory failure

Return to factory/service center.

 

Battery back-up failure.

Return to factory/service center.

 

 

 

Fault displayed in sensor location

Sensor missing.

Properly install sensor.

 

Invalid sensor readings.

Check sensor calibration. Replace if

 

 

necessary.

 

 

 

30

Page 33
Image 33
O2 Innovations pmn manual Troubleshooting, TARGET Series, ENMET Corporation, Symptom, Possible Cause, Action