TARGET Series | ENMET Corporation |
10.0 Troubleshooting
Symptom
Possible Cause
Action
Unit will not turn on. | Dead battery | Plug instrument into charger. Check |
|
| charge status. |
|
|
|
Sensor fails calibration | Failed sensor | Replace sensor and recalibrate |
| Wrong gas concentration | Use proper calibration gas. |
| Old calibration gas | Replace calibration gas. |
|
|
|
Calibration due on status line | Sensor hasn’t been calibrated in last 30 | Check Cal Status in |
| days. | Calibrate affected sensor(s). |
|
|
|
Cal error on status line | Sensor failed a calibration. | Check Cal Status in |
|
| Calibrate affected sensor(s). Replace |
|
| sensor if necessary and recalibrate. |
|
|
|
Setup/config error on status line | Memory failure | Return to factory/service center. |
| Battery | Return to factory/service center. |
|
|
|
Fault displayed in sensor location | Sensor missing. | Properly install sensor. |
| Invalid sensor readings. | Check sensor calibration. Replace if |
|
| necessary. |
|
|
|
30