8.0 Default Alarm and Configurations

TARGET Series

 

 

 

 

 

ENMET Corporation

8.0 Default Alarm and Configurations

 

 

 

 

 

 

Table 3: Default Alarm

 

 

 

 

 

 

 

 

 

 

 

Gas

Full Scale

Resolution

Alarm 1

Alarm 2

TWA

STEL

Calibration

 

Display

 

 

 

 

 

Gas Value

Ammonia

99 ppm

1 ppm

25 ppm

50 ppm

25 ppm

35 ppm

25 ppm NH3

Arsine

1.00 ppm

0.01 ppm

0.05 ppm

0.10 ppm

0.10 ppm

0.10 ppm

0.5 ppm PH3

 

 

 

 

 

 

 

0.5 ppm /AsH3

BRH

200 ppm

1 ppm

100 ppm

150 ppm

X

X

100 ppm C7H8

 

 

 

 

 

 

 

(toluene)

Carbon monoxide

500 ppm

1 ppm

35 ppm

200 ppm

35 ppm

200 ppm

100 ppm CO

 

 

 

 

 

 

 

 

Δ Chlorine

9.9 ppm

0.1 ppm

0.5 ppm

5.0 ppm

.05 ppm

1.0 ppm

5 ppm Cl2

Chlorine dioxide

3.00 ppm

0.01 ppm

0.10 ppm

0.30 ppm

0.10 ppm

0.30 ppm

1.0 ppm Cl2

CO+ (BRH)

100 ppm

1 ppm

35 ppm

75 ppm

X

X

35 ppm CO

 

 

 

 

 

 

 

 

Combustible

99 %LEL

1 %LEL

10 %LEL

50 %LEL

X

X

50 %LEL CH4

HS+ (BRH)

40 ppm

1 ppm

10 ppm

20 ppm

X

X

20 ppm H2S =

 

 

 

 

 

 

 

50 ppm CO

 

 

 

 

 

 

 

 

Δ Hydrogen chloride

20.0 ppm

0.1 ppm

5.0 ppm

10.0 ppm

5.0 ppm

5.0 ppm

10 ppm HCl

Hydrogen cyanide

20.0 ppm

0.1 ppm

4.7 ppm

10.0 ppm

4.7 ppm

4.7 ppm

10 ppm HCN

 

 

 

 

 

 

 

 

Hydrogen fluoride

10.0 ppm

0.1 ppm

0.3 ppm

6.0 ppm

3.0 ppm

3.0 ppm

5.0 ppm HF

 

 

 

 

 

 

 

 

Hydrogen sulfide

200 ppm

1 ppm

10 ppm

20 ppm

10 ppm

15 ppm

20 ppm H2S

Ozone

3.00 ppm

0.01 ppm

0.10 ppm

0.30 ppm

0.10 ppm

0.10 ppm

0.3 ppm O3

Oxygen

30 %volume

0.1 % volume

19.5%

23.5%

X

X

20.9 % O2

Phosphine

2.00 ppm

0.01 ppm

0.30 ppm

1.50 ppm

0.30 ppm

1.5 ppm

0.5 ppm PH3

Nitrogen dioxide

30.0 ppm

0.1 ppm

1.0 ppm

5.0 ppm

1.0 ppm

1.0 ppm

5 ppm NO2

Sulfur dioxide

30.0 ppm

0.1 ppm

2.0 ppm

10.0 ppm

2.0 ppm

5.0 ppm

10 ppm SO2

Δ Ethylene oxide

10 ppm

0.1 ppm

5.0 ppm

7.5 ppm

5.0 ppm

5.0 ppm

5 ppm ETO

X = not applicable.

ΔThese gases require the Sampling Shield p/n 04836-000 to be in place and pump activated for accurate detection and calibration.

Sensor response time with internal pump, if the tubing is 20 feet long, allow at least 10 seconds for the test sample to make it through the tubing. The sensor response time is typically 30 – 60 seconds to achieve 90% o f the final reading, depending on the specific sensor and the gas being detected. Therefore, the total 'system' response time for an instrument using the internal sampling pump and 20 feet of hose is approximately 40-70 seconds to achieve 90% of the final reading.

 

Table 4: Default Configuration

Operation Menu:

Basic

 

 

Confidence chirp interval:

120 seconds

 

 

Backlight duration:

30 seconds

 

 

Combustible gas display:

Methane

 

 

Datalogging:

Active (requires TARGETTrakker to retrieve)

 

 

Datalogging interval:

1 per minute

 

 

TWA/STEL Alarms:

Active

 

 

Locations:

Enabled (requires TARGETTrakker to set)

 

 