O2 Innovations pmn Various Calibration Adapter Attachments, TARGET Series, ENMET Corporation

Models: pmn

1 36
Download 36 pages 1.53 Kb
Page 17
Image 17
Figure 8: Various Calibration Adapter Attachments

TARGET Series

ENMET Corporation

Sensor end cap w/cal shield

Regulator

Humidifier Bowl

Cylinder,

Calibration Gas

Example of Storage

Calibration Adapter, BRH 04834-005Case used as Support for Humidifier

Sensor end cap w/cal shield

Sensor end cap

w/cal shield

Humidifier Bowl

Regulator

Regulator

 

 

Top View of Humidifier Bowl

To Sensor

To Regulator

 

Cylinder, Calibration Gas

Calibration Adapter, BRH

04834-002

Cylinder, Calibration Gas

Calibration Adapter, Reactive Gases

Figure 8: Various Calibration Adapter Attachments

Apply the span gas and press the SELECT button to initiate the sample count down timer. Count down is shown on the lowest line of the display, at the end of the count down the span reading is stored in memory. Remove the span gas and the calibration shield. Calibration of this sensor is complete.

If the sensor is within acceptable sensitivity parameters, OK is displayed next to the gas momentarily before the display returns to the Calibration menu.

If there is a problem, Cal error is displayed next to the appropriate channel. See section 10.0 for troubleshooting.

CAL: O2/CO/H2S/Mth Sampling MULTI-BLEND MENU:Abort

Sample in 120 sec.

Example:

Display during and after Calibration

O2:

OK

CO:

Cal error

H2S:

OK

Mth:

OK

 

 

Repeat the above procedure for each individual sensor as required. If a calibration error occurs it will also be displayed as a general condition on the Status Line of the Main Gas Display and can be viewed in the Cal Status screen of the Pop-Up menu.

Exit the Calibration menu when completed. Remove span gas and calibration shield. Then exit the Standard Maintenance menu.

14

Page 17
Image 17
O2 Innovations pmn Various Calibration Adapter Attachments, Example Display during and after Calibration, TARGET Series