PART 7
APPROVALS INFORMATION
7.1CE APPROVALS INFORMATION
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Double Insulation |
| |
Pollution Degree 2 |
| |
Dielectric withstand Test per 1 min |
| |
• | Power to Input/Output: | 2300 Vac (3250 Vdc) |
• | Power to Input/Output: | 1500 Vac (2120 Vdc) |
|
| (Low Voltage dc Power Option*) |
• | Power to Relays/SSR Output: | 2300 Vac (3250 Vdc) |
• | Ethernet to Inputs: | 1500 Vac (2120 Vdc) |
• | Isolated RS232 to Inputs: | 500 Vac (720 Vdc) |
• | Isolated Analog to Inputs: | 500 Vac (720 Vdc) |
• | Analog/Pulse to Inputs: | No Isolation |
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum
Transients Overvoltage Surge (1.2 / 50uS pulse)
• | Input Power: | 2500 V |
• | Input Power: | 1500 V |
|
| (Low Voltage dc Power Option*) |
• | Ethernet: | 1500 V |
• | Input/Output Signals: | 500 V |
Note: *Units configured for external low power dc voltage,
EMC:
EN 61326:2006, Class B Emissions EN 61326:2006, Immunity
EN 55022:2006 + A1:2007, Class B, (CISPR 22 + A1:2005) - Emissions Ethernet EN 55024:1998 + A1:2001 + A2:2003 - Immunity Ethernet
EN
Note: **I/O signal and control lines require shielded cables and these cables must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.
7.2FCC
This device complies with Part 15, Subpart B, Class B of the FCC rules.