N

-250 ~ 1300 ˚C

±1 ˚C

2.2MΩ

 

(-418 ~ 2372˚F)

 

 

L

-200 ~ 900 ˚C

±1 ˚C

2.2MΩ

 

(-328 ~ 1652 ˚F)

 

 

PT100

-210 ~ 700 ˚C

±0.4 ˚C

1.3KΩ

(DIN)

(-346 ~ 1292 ˚F)

 

 

PT100

-200 ~ 600 ˚C

±0.4 ˚C

1.3KΩ

(JIS)

(-328 ~ 1112 ˚F)

 

 

mV

-8 ~ 70mV

±0.05%

2.2MΩ

mA

-3

~ 27mA

±0.05%

70.5Ω

0~1V

-0.12 ~ 1.15V

±0.05%

332KΩ

0~5V

-1.3

~ 11.5V

±0.05%

332KΩ

1~5V

-1.3

~ 11.5V

±0.05%

332KΩ

0~10V

-1.3

~ 11.5V

±0.05%

332KΩ

Digital Input Card (DI181)

Channels: 6 per card

Logic Low: -30V minimum, 0.8V maximum

Logic High: 2V minimum, 30V maximum

External pull-down Resistance: 1KΩ maximum

External pull-up Resistance: 1.5MΩ minimum

Digital Output Card (DO181) Channels: 6 per card Contact Form: N.O. (form A)

Relay Rating: 5A/240 VAC, life cycles 200,000 for resistive load

24VDC Auxiliary Power Supply Card ( AP181) Channels: to be used for 6 transmitters

Output Rating: 24 ± 1 VDC, 180mA in maximum, 30mA / each channel

COMM Module (CM181)

Interface: RS-232 (1 unit), RS-485 or RS-422 (up to 247 units)

Protocol: Modbus Protocol RTU mode

Address: 1-247

Baud Rate: 0.3~38.4 Kbits/sec.

Measured data Bits: 7 or 8 bits

Parity Bit: None, Even or Odd

Stop Bit: 1 or 2 bits

Standard Ethernet Communication

Protocol: Modbus TCP/IP, 10 Base T

Ports: AUI (Attachment Unit Interface) and RJ-45, Auto- detect capability

11

Page 11
Image 11
Omega Engineering RD8900 user manual Digital Input Card DI181, Comm Module CM181, Standard Ethernet Communication