
CTS 710 SONET Test Set User Manual 3–121
Setting Tributary Receive ParametersThis section describes how to manually set the receive rate, receive
level, framing, and test pattern for tributary signals.
Pressing AUTOSCAN sets the receive parameters automatically
based on the received signal.
Steps for Setting Parameters
The sequence of steps for setting receive parameters varies with
signal to be received. Figure 3–44 shows the sequence of steps
required to set the receive parameters for tributary signals (for
example, DS3). Figure 3–45 shows the sequence of steps required to
set the receive parameters for demultiplexing a DS1 signal from a
DS3 signal.
Set
Rate Set Level Set
Framing Set Test
Pattern
Figure 3–44: Sequence for Setting Tributary Signal Parameters
Set
Rate Set Level Set DS3
Framing Set DS1
Under Test Set Test
Pattern
Set DS1
Framing
Figure 3–45: Sequence for Setting Signal Parameters to Demultiplex a DS1 from a
DS3
CTS 710 SONET Test Set User Manual 3–121
Setting Tributary Receive ParametersThis section describes how to manually set the receive rate, receive
level, framing, and test pattern for tributary signals.
Pressing AUTOSCAN sets the receive parameters automatically
based on the received signal.
Steps for Setting Parameters
The sequence of steps for setting receive parameters varies with
signal to be received. Figure 3–44 shows the sequence of steps
required to set the receive parameters for tributary signals (for
example, DS3). Figure 3–45 shows the sequence of steps required to
set the receive parameters for demultiplexing a DS1 signal from a
DS3 signal.
Set
Rate Set Level Set
Framing Set Test
Pattern
Figure 3–44: Sequence for Setting Tributary Signal Parameters
Set
Rate Set Level Set DS3
Framing Set DS1
Under Test Set Test
Pattern
Set DS1
Framing
Figure 3–45: Sequence for Setting Signal Parameters to Demultiplex a DS1 from a
DS3