Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
Table 7-7-9: Enabling CAL-ON-NO2
Step Number Action Comment
1. Press SETUP-
MORE-VARS
Causes the M200A go to the VARS(Variables) menu
2. Enter the VARS
password, then
ENTR
Password value is 8 1 8
3. Scroll the VARS
menu to select the
CAL-ON-NO2
variable
By pressing NEXT or PREV
4. Press EDIT Once you have entered EDIT, set the value to ON
5. Press ENTR ENTR will enable the CAL-ON-NO2 feature
6. Press EXIT Each EXIT press returns the M200A to the next higher menu
level. The highest level is the SAMPLE mode
Once CAL-ON-NO2 is enabled the procedure for zeroing and spanning the instrument is
identical to that given in Section 7.2 for Zero/Span Valves.