Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
7.8.5.1 Zero Calibration Procedure
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero
value. The following Table 7-7-14 details the zero calibration procedure.
Table 7-7-14: Zero Calibration Procedure
Step Number Action Comment
1. Press CAL The M200A enters the calibrate mode from sample mode.
NOTE:
The analyzer does not operate the zero/span valves in this mode,
the zero gas enters through the sample port.
2. Wait 10 min Wait for reading to stabilize at the zero value.
3. Press ZERO If you change your mind after pressing ZERO, you can still
press EXIT here without zeroing the instrument.
4. Press ENTR Pressing ENTR actually changes the calculation equations.
5. Press EXIT M200A returns to the SAMPLE mode.
7.8.5.2 NO/NOx Calibration Procedure
Adjust the NO concentration to approximately 80% of the URL of the NO channel. The expected
NO and NOx span concentrations can be determined by measuring the cylinder and diluent flows
and computing the resulting concentrations. If there is any NO2 impurity in the NO standard gas
it should be taken into account when the NOx concentration is entered during the NO/NOx
channel calibration. This is done by ADDING the impurity concentration to the NO
concentration to get the NOx concentration for calibration. Calculate the exact NO and NOx
concentrations as follows:
F
]
[NO x
F
=
]
[NO
T
STD
NO
OUT Equation 7-8Enter the respective concentrations using the
procedure in Table 7-13. The expected span concentrations need not be re-entered each time a
calibration is performed unless they are changed.
Enter the expected NOx and NO span gas concentrations:
7-22
PRINTED DOCUMENTS ARE UNCONTROLLED