Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-14: RS-232 Command Summary
Commands Definition
? [id] Print help screen. ID is an optional instrument ID number
T [id] LIST Print all active test messages
T [id] LIST name or
T [id] name
Print single test message "name" from Table 5-15
W [id] LIST Print all active warnings
W [id] CLEAR name or
W [id] name
Clear single warning message "name" from Table 5-16
C [id] command Execute calibration "command" from Table 5-17
D [id] LIST Prints all I/O signal values
D [id] name Print single I/O signal value/state
D [id] name=value Sets I/O signal to new "value"
D [id] LIST NAMES Lists diagnostic test names
D [id] ENTER name Enters and starts 'name' diagnostic test
D [id] EXIT Exits diagnostic mode
D [id] RESET Resets analyzer(same as power-on)
D [id] RESET RAM Sy stem reset, plus erases RAM. Initializes DAS, NO, NOx, NO2 conc
readings, calib not affected.
D [id] RESET EEPROM System reset, plus erases EEPROM (RESET RAM actions + setup
variables, calibration to default values). Restores all factory defaults.
D [id] PRINT Prints properties for all data channels (DAS)
D [id] PRINT "name” Prints properties for single data channel. Quotes around name are
required.
D [id] REPORT "name"
[RECORDS=number]
[COMPACT|VERBOSE]
Prints DAS records for a data channel. Quotes around name are
required. Parameters in brackets are optional.
V [id] LIST Print all setup variable names and values
V [id] name Print individual setup variable value
V [id] name=value Sets setup variable to new "value"
V [id] CONFIG Print analyzer configuration
V [id] MODE Print current analyzer mode
5-29 PRINTED DOCUMENTS ARE UNCONTROLLED