Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
Table 9-2: Front Panel Warning Messages (Continued)
Message Description
PMT TEMP WARNING PMT temp. out of spec. The PMT temp has its own
proportional controller on the preamp (01141) board.
Warnings will occur during initial operation. The warning can
occur if the 7 pin connector to the interior of the sensor is not
plugged in. The power connector to the PSM should be
checked for proper voltage (+15 VDC ± 0.5).
MOLY TEMP WARNING Molycon temp. out of spec. The Moly temp is controlled by
the CPU. It has a thermocouple with amplifier on the 01086
Status/Temp board. Because of the high temperature of the
Moly (315 C), the moly temp warning will tend to be the last
warning to clear as the instrument is powered on.
CANNOT DYN ZERO Dynamic zero cal. out of spec. The reading of the PMT was
too high for the ZERO button to appear. Make sure the
instrument is receiving zero gas. Check for dirty reaction cell.
Do the factory calibration procedure located in Section 9.1.6.
CANNOT DYN SPAN Dynamic span cal. out of spec. The reading of the PMT was
too high or low for the SPAN button to appear. Make sure the
instrument is receiving correct concentration span gas. Make
sure the expected span concentration is entered. Check for
dirty reaction cell. Do the factory calibration procedure located
in Section 9.1.6.
OZONE GEN OFF Ozone Generator is off. See Table 9-9-17for conditions.
AZERO WRN XXX.X MV The AutoZero circuit compensates for detector dark current,
and background light. Certain electrical faults cause high
readings to be added to the filter. First, the cause of the high
ZutoZero reading should be found and repaired, then wait 15
minutes for the AutoZero filter to clear itself out.
A/D NOT INSTALLED V/F (00514) board has failed. The V/F board did not respond
to commands from the CPU. This probably means 1. board not
seated in socket 2. defective board 3. defective back plane
connector.
HVPS WARNING High Voltage Power Supply voltage out of limits. Limits are
400-900 VDC.
DCPS WARNING DC Power Supply voltage out of limits. Limits are 2500
± 500 mV.
9-10 PRINTED DOCUMENTS ARE UNCONTROLLED