Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
Table 5-5-11: RS-232 Port Setup - Front Panel
Decimal Value Description
1 Turns on quiet mode (status messages suppressed)
2 Places analyzer in computer mode (no echo of chars)
4 Enables Security Features (Logon, Logoff)
8 Enables TELEDYNE API protocol and setup menus
16 Enable alternate protocol
32 Enable multidrop protocol
64 Enable modem setup string
128 Ignore RS-232 line errors
4096 Enable command prompt
NOTE
To enter the correct value, ADD the decimal values of the features you
want to enable. For example if LOGON and front panel RS-232
menus are desired, the value entered would be 4 + 8 = 12.
Port Communication
The RS-232 can be set up in interactive or computer mode of communication. If the port is
attached to a computer running an instrument interface program, it needs to have different
characteristics than if used interactively by a human operator.
When an operator is communicating with the analyzer via a terminal, the analyzer should be
placed into TERMINAL MODE, which echoes keystrokes, allows editing of the command line
using the backspace and escape keys, and allows recall of the previous command. When a host
computer or data logger is connected to the analyzer, it should be placed into COMPUTER
MODE, which does not echo characters received or allow the special editing keys. See Table 6-
6-5 for relevant commands.