Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247

Table 7-7-21: Calibration Quality Check

Step Number Action Comment
1. Scroll the TEST
function menu until
the NOx SLOPE is
displayed
The SLOPE value for NOx should be 1.0 ± 0.3. If the value is
not in this range, check Section 9.1.6. If the SLOPE value is in
the acceptable range the instrument will perform optimally.
2. Scroll the TEST
function menu until
the NO SLOPE is
displayed
The SLOPE value for NO should be 1.0 ± 0.3. If the SLOPE is
in the acceptable range the instrument will perform optimally. If
the value is not in this range, check Section 9.1.6.
NOTE:
The NO and NOx slopes should be equal within ± 0.1.
3. Scroll the TEST
function menu until
the NOx OFFSET is
displayed
This number should be near zero. A value of 0.0 – 20 + 150
indicates calibration in the optimal range. This number already
has the AutoZero value subtracted out and is mainly the
background signal due to the molybdenum converter. If the
OFFSET value is outside this range, check Section 9.1.6.
4. Scroll the TEST
function menu until
the NO OFFSET is
displayed
The instrument will now display the NO OFFSET value. It
should also have a value near zero (0.0 –20 + 150 mV). This
number already has the AutoZero reading subtracted out and
should be near zero. If the OFFSET value is outside this range,
check Section 9.1.6.
After the above procedure is complete, the M200A is ready to measure sample gas.

7-36

PRINTED DOCUMENTS ARE UNCONTROLLED