Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
3.2 EPA Equivalency Designation
Advanced Pollution Instrumentation, Inc., Model 200A Nitrogen Oxides Analyzer is designated
as Reference Method Number RFNA-1194-099 as defined in 40 CFR Part 53, when operated
under the following conditions:
1. Range: Any range from 50 parts per billion (ppb) to 1 ppm.
2. Ambient temperature range of 5 to 40oC.
3. Line voltage range of 105-125 VAC, 60Hz; 220-240 VAC, 50Hz.
4. With 5-micron TFE filter element installed in the internal filter assembly.
5. Sample flow of 500 ± 50 cc/min.
6. Vacuum pump (internal or external) capable of 10"Hg Abs pressure @ 2 slpm or better.
7. Software settings:
A. Dynamic span OFF
B. Dynamic zero OFF
C. Cal-on-NO2 OFF
D. Dilution factor OFF
E. AutoCal ON or OFF
F. Independent range ON or OFF
G. Autorange ON or OFF
H. Temp/Pres compensation ON
I. Converter Eff. Acceptable values of 0.96 to 1.02
Under the designation, the Analyzer may be operated with or without the following options:
1. Rack mount with slides.
2. Rack mount without slides, ears only.
3. Rack mount for external pump w/o tray.
4. Stainless steel zero/span valves.
5. Internal zero/span.
6. NO2 permeation tube - uncertified 0.4ppm @ 0.7 lpm.
7. NO2 permeation tube - certified 0.4ppm @ 0.7 lpm.
8. NO2 permeation tube - certified 0.8ppm @ 0.7 lpm.
9. NO2 permeation tube - uncertified 0.8ppm @ 0.7 lpm.
10. 4-20mA, isolated outputs.
11. Status outputs.
12. RS-232 output.
13. Internal pump or external pump.
14. 1 micron sample filter.