Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247

3 SPECIFICATIONS, AGENCY APPROVALS, WARRANTY

3.1 Specifications

Ranges In 1ppb increments from 50ppb to 20,000ppb
Range Modes Single, Independent, AutoRange
Noise at Zero1 0.2 ppb RMS
Noise at Span1 <0.5% of reading above 50 ppb
Lower Detectable Limit2 0.4 ppb RMS
Zero Drift3 <0.5 ppb/24 hours
Zero Drift 1 ppb/7 days
Span Drift <0.5% FS/7 days
Lag Time 20 sec
Rise Time 95% in <60 sec
Fall Time 95% in <60 sec
Sample Flow Rate 500 cc/min. ± 10%
Linearity 1% of full scale
Precision 0.5% of reading
Temperature Range 5-40oC
Temp Coefficient < 0.1% per oC
Humidity 0-95% RH non-condensing
Voltage Coefficient < 0.1% per V
Dimensions HxWxD 7" x 17" x 23.6" (18cm x 43cm x 61cm)
Weight, Analyzer 43 lbs (20 kg) w/external pump
Weight, Analyzer 55 lbs (25 kg) w/internal pump
Weight, Ext Pump Pack 16 lbs (7 kg)
Weight, Internal Pump 5 lbs (2 kg)
Power, Analyzer 100 V ~ 50/60 Hz, 120V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 125 watts
Power, Analyzer5 230 V ~ 50 Hz, 2.5A peak
Power, Ext Pump 110 V ~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 295 watts
Power, Ext Pump5 230 V ~ 50 Hz, 2.5A peak
Power, Int Pump 110 v/50/60 Hz, 60 watts
Environmental Installation Category (Over-voltage Category) II Pollution Degree 2
Recorder Output4 0-100 mV, 0-1, 5, 10v
Analog Resolution 1 part in 1024 of selected voltage or current range
Status Option 12 Status Outputs from opto-isolator
Measurement Units ppb, ppm, ug/m3, mg/m3

1. As defined by USEPA.
2. Defined as twice the zero noise level.
3. At constant temperature and voltage.
4. Bi-polar.
5. Electrical rating for CE Mark compliance.

3-1

PRINTED DOCUMENTS ARE UNCONTROLLED