Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 2-12 PRINTED DOCUMENTS ARE UNCONTROLLED
Table 2-2-6: Final Test and Calibration Values
Test Values Observed
Value Units Nominal Range Reference Section
RANGE PPB 50-20,000 5.3.4
STABILITY
(Zero Gas)
PPB < 0.2 9.1.1, Table 9-9-1,
9.2.5
SAMP FLW cc/min 500 ± 50 9.3.7, Table 9-9-1
OZONE FL cc/min 60 - 90 9.3.6, 9.3.7
PMT mV 0-5000 9.3.8.1, Table 9-9-1
AZERO mV -20 - 150 Table 9-9-1
HVPS V 450 - 900 constant 9.3.8.5
DCPS mV 2500 ± 200 9.3.5, 9.3.4
RCELL TEMP oC 50 ±1 9.3.8.2
BOX TEMP oC 8-48 9.3.4.1
PMT TEMP oC 7 ± 1 9.3.8.4
IZS TEMP oC 50 ± 0.4 9.3.9
MOLY TEMP oC 315 ± 5 9.3.4.1
RCEL PRES IN-Hg-A 4 - 10 constant 9.3.7
SAMP PRES IN-Hg-A 25 - 30 constant 9.3.7
Electric Test & Optic Test
Electric Test
PMT Volts mV 2000 ± 500 9.1.3.2
NO Conc PPB 1000 ± 250 9.1.3.2
NOx Conc PPB 1000 ± 250 9.1.3.2
Optic Test
PMT Volts mV 2000 ± 1000 9.1.3.3
NO Conc PPB 1000 ± 500 9.1.3.3
NOx Conc PPB 1000 ± 500 9.1.3.3
(table continued)