2.2 Front Panel Display

2.2 Front Panel Display

This section describes the operator interface from the point of view of the front panel. The front panel consists of a 2-line by 40-character alphanumeric display, 8 pushbuttons, and 3 status LED’s. Each of these features is described below.

2.2.1 Front panel display fields

The display is divided into 4 main "fields": the Mode field in the upper left, the Message/Test Function field in the top center, the ozone Concentration field consisting of the most recent instantaneous ozone value field in the upper right, and the Menu field which occupies the entire bottom line of the display. The Menu field is used to define the function of the 8 pushbuttons directly below the display. The buttons are then used for selecting menu items and are also used for entering values such as alarm levels. A typical display is shown in Figure 2.1.

MODE

 

MESSAGE/TEST

CONCENTRATION

FIELD

 

MEASUREMENT FIELD

 

FIELD

SAMPLE

TIME=15:58:21

 

 

12.0 wt%

<TST

TST>

ALRM

MSG

CLR

SETUP

 

 

BUTTON DEFINITIONS

 

FIGURE 2.1 - MODEL 450H FRONT PANEL

Mode Field

The mode field indicates the current mode of the Monitor. Table 2.1 lists all the possible modes in the Monitor and their meanings.

TABLE 2.1 SYSTEM MODES

Mode

 

Meaning

SAMPLE

 

Monitoring

ALARM STATUS

 

Display of Alarm Status

SETUP xxx (1)

Configuring monitor (monitoring continues)

DIAG CAL

 

Calibration Menu

DIAG D/A

Configure and Calibrate Digital to Analog converters

DIAG AOUT

 

Test analog output

DIAG CFG

 

Instrument Configuation List

 

(1) xxx = software

revision (e.g. A.9)