1.4 Specifications
Measurement Principle | UV Absorption (Beer Lambert Law) |
Ranges | |
| |
Measurement Units | wt%, g/m3, g/Nm3 (Operator selectable) |
Precision/Repeatability | ±0.5% of full scale range |
Resolution | 0.1 % w/w |
Linearity | 1% |
Response Time (95%) | <5 sec to 95% |
Compensation | Pressure, temperature, gas molecular weight |
Gas Flow Rate | 0.5 - 2 LPM |
Temperature Range | |
Alarm Relay Mode | Latching or |
19” Rack Mount Enclosure | 7" x 17" x 18.2" |
Dimensions (H x W x D) | (178 mm x 432 mm x 463 mm) |
NEMA 4X Enclosure | 19.50” x 17.25” x 9.63” |
Dimensions (H x W x D) | (495 mm x 438 mm x 245 mm) |
Weight | 28 lb. (12.7 kg) - 19” Rack Mount Enclosure |
| 33 lb. (15 kg) - NEMA 4X Enclosure |
Power | 110V/60 Hz, 220V/50 Hz, 240V/50Hz 250 watts |
| 230V~, 50Hz, 2.5A |
Environmental Conditions | Installation Category (Overvoltage Category) II |
| Pollution Degree 2 |
Maximum Operating Altitude | 2000 meters |
Analog Output Voltage Mode | 100mV, 1V, 5V, 10V (User Selectable) |
Isolated Analog Output | Maximum voltage between outputs and ground |
Mode* | 60V peak |
Degree of Protection (IP Code) | IPX0 (450 Rack Mount) |
*Optional | IPX66 (450 Nema) |
|
1.5 Installation (19” Rack Mount Enclosure)
The Model 450H 19” Rack version is shipped with the following standard equipment:
P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 11