1.Instruction manual.
2.Power Cord
CAUTION
To avoid personal injury, always use two persons to lift and
carry the Model 450H.
Upon receiving the Model 450H please do the following:
1.Verify no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.)
2.Before connecting power to the instrument, check the serial number tag to verify that the instrument has been configured for the correct line voltage and frequency connections.
CAUTION
Verify that the instrument is set up for proper line voltage and
frequency.
3.Connect an exhaust line to the fitting labeled ‘Exhaust.’ If the M450H has been configured with the ozone killer option, this line should be vented to an outside area, since the exhaust gas may still contain trace levels of ozone. If the M450H has not been equipped with an ozone killer, this line should be vented to an ozone destruct unit or other appropriate device.
CAUTION
Exhaust gas from the M450H may contain dangerous levels of
ozone!
4.Connect the ozone delivery line to the ¼” inlet fitting on the filter assembly on the rear panel. The ozone delivery pressure should be regulated to no more than 30psig. All tubing used should be made of ozone resistant material such as PTFE(Teflon™) or PFA. If new, unconditioned tubing is used for the ozone delivery line, the ozone concentration may be affected until the tubing becomes conditioned.
P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 12