Output

 

Output

 

 

Description

 

 

Normal Operating State

 

 

 

 

 

 

 

1

 

Power OK

ON

2

 

 

Diagnostics Mode(Instrument not

 

OFF

 

 

 

 

monitoring)

 

 

 

3

 

 

Temperature Fault

OFF

4

 

 

Pressure Fault

OFF

5

 

 

UV Lamp Fault

OFF

6

 

 

Flow Fault

OFF

7

 

 

Performing Auto-Zero Calibration

OFF

8-11

 

Not Used

N/A

12

 

 

System OK(no faults)

ON

TABLE 1.1 DIGITAL OUTPUTS

Each digital output is configured as a Collector/Emitter pair. The labels are ‘C’ and ‘E’ for these contacts, respectively. Figure 1.2 below shows the most common way of connecting the digital outputs to an external device such as PLC. Note: Most devices, such as PLC’s, have internal provision for limiting the current that the input will draw from an external device. When connecting to a unit that does not have this feature, external dropping resistors must be used to limit the current through the transistor output to 50mA or less.

FIGURE 1.5 - CONNECTING DIGITAL OUTPUTS

1.6.4 Control Inputs

There are 6 digital control inputs on the rear panel; they are labeled ‘CONTROL IN’ (See Figure 1.4.) The control inputs are used for remote control of the M450H by a device such as a PLC. These inputs are triggered by providing a contact closure or low impedance current path between the + and ground contacts. This can be done by using a mechanical switch or isolated transistor type output from another device, such as a PLC. Never connect a voltage level output from another device to these contacts. The functions of the control inputs is summarized below in Table 1.2:

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 20

Page 20
Image 20
Teledyne 450H instruction manual Control Inputs, Output, Description, Normal Operating State, 1 DIGITAL OUTPUTS