22-11

Using the EFI Spectrometer ES-1000

The following illustration shows the different parts of the ES-1000.

Measure button

Sample aperture

White tile reference

Calibration cradle

Calibrating the GA-1121 with the ES-1000

You can use ColorWise Pro Tools in conjunction with the ES-1000 to calibrate the GA-1121.

TO CALIBRATE THE GA-1121USING THE ES-1000

1.Start Calibrator.

For instructions on starting Calibrator, see page 2-6.

2.Choose EFI Spectrometer ES-1000 as the measurement method.

3.In the Check Print Settings pane, choose the desired calibration set.

Choose the appropriate calibration set for the mode and type of media which are needed for calibration.

NOTE: For this calibration to take effect, the calibration set must be associated with one or more output profiles. There is no need to make any new associations for the default calibration set, since it is already associated with the default output profile.

4.In the Generate Measurement Page pane, click Print.

The Print Options dialog box appears.

Page 47
Image 47
Toshiba manual Calibrating the GA-1121 with the ES-1000, Using the EFI Spectrometer ES-1000, Start Calibrator