22-14 Color Calibration

14.Repeat step 10 through step 13 until all the strips have been measured.

NOTE: You must take the measurements in the order outlined in the Status field.

15.When all the patches have been read successfully, click Accept.

16.Click Apply in the Calibrator window to implement the new calibration measurement.

17.Click OK in the Information dialog box.

This completes the GA-1121 calibration process.

Using the EFI Densitometer ED-100

The EFI Densitometer ED-100 is a hand-held instrument that measures the density of printed ink and toner. Like the DTP32, the ED-100 is a reflection densitometer that you can use to calibrate the GA-1121.

Setting up the ED-100

To use the ED-100 for calibration purposes, you must first set it up for use with your computer. For complete instructions on setting up the ED-100, see the documentation that is included with the instrument.

The following illustration shows the components of the ED-100.

Indicator light

Measure button

Sample aperture (underneath)

Page 50
Image 50
Toshiba GA-1121 manual Using the EFI Densitometer ED-100, Setting up the ED-100