Xantrex Technology GPIB-XPD manual Voltage Mode Calibration

Models: GPIB-XPD

1 58
Download 58 pages 4.56 Kb
Page 53
Image 53

Calibration

Voltage Mode Calibration

3.Send command VRLO; IRLO to the power supply. Wait for the supply to settle. Measure and record the output shown on the external voltmeter. Send VRLO again.

4.Send VRHI; IRHI to the supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send VRHI again.

5.Send the command VRDAT <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter after the VRLO and VRHI commands were sent. The processor calculates the offset value required to calibrate the power supply.

When the power supply is calibrated, the low to high voltage readback calibration values (offsets) are stored as constants.

6.Use commands VSET and VOUT? commands to confirm that the calibration was successful and that linearity is observed. Refer to the voltage readback accuracy specification in Section 1.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.