Cleaning the imaging guides — thorough cleaning
Thorough cleaning of the imaging guides is recommended at the start of the day and beginning of each shift.
NOTE: Hold the imaging guides at the green tabs to avoid getting fingerprints on the imaging guides during the cleaning procedure.
Upper imaging guide
1.Open the pod.
2.Turn the screw on each end of the upper imaging guide and remove it from its position.
3.Clean the imaging guide thoroughly with a Staticide wipe.
4.Wipe the imaging guide again with an
IMPORTANT: There is no cover glass between the upper lamps. Do not insert any objects into the area between the upper lamps, as you may damage the reflective mirrors in the scan module.
5. Reinstall the upper imaging guide.