Xantrex Technology XPF 35-10 manual Output Specifications

Page 4

Specification

OUTPUT SPECIFICATIONS

Voltage Range:

0V to 35V

Current Range:

0A to 10A

Power Range:

Up to 175W

Output Voltage Setting:

By coarse and fine controls.

Output Current Setting:

By single logarithmic control.

Operating Mode:

Constant voltage or constant current with automatic cross-over

 

provided that the power demanded stays within the power

 

envelope, see graph. Outside of this envelope the output becomes

 

unregulated.

Output Switch:

Electronic. Preset voltage and current displayed when off.

Output Terminals:

4mm terminals on 19mm (0·75”) pitch. 15A max.

Sensing:

Switchable between local and remote. Spring-loaded push

 

terminals for remote connection.

Output Impedance:

Typically <5min constant voltage mode.

 

Typically >5kin constant current mode (voltage limit at max).

Output Protection:

Forward protection by Over-Voltage Protection (OVP) trip;

 

maximum voltage that should be applied to the terminals is 50V.

 

Reverse protection by diode clamp for reverse currents up to 3A.

OVP Range:

10% to 110% of maximum output voltage set by front panel

 

screwdriver adjustment.

Line & Load Regulation:

<0.01% of maximum output for a 10% line change; <0.05% of

 

maximum output for a 90% load change.

Ripple & Noise

5mVrms max; typically <2mVrms, <20mV pk-pk, both outputs fully

(20MHz bandwidth):

loaded (7A @ 25V), CV mode.

Transient Load Response:

<2ms to within 100mV of set level for 90% load change.

Temperature Coefficient:

Typically <100ppm/°C

Status Indication:

Output on lamp.

 

Constant voltage mode lamp.

 

Constant current mode lamp.

 

Unregulated (power limit) lamp

 

Trip message on display.

 

 

 

3

Image 4
Contents XPF Series Dual Output 35V 10A Powerflex DC Power Supply XPFPart Number Contact InformationTable of Contents Specification Output SpecificationsMeter Specifications Safety Emissions ImmunityInstallation Mains Operating VoltageMains Lead MountingConnections Operation Connection to the Load Remote SensingSeries or Parallel connection with other units ProtectionMaintenance CleaningVentilation FuseCalibration Access to Calibration AdjustmentsEquipment Required Set Power Limit Voltage CalibrationCurrent Calibration OVP CalibrationSécurité AVERTISSEMENT! CET Instrument Doit Etre Relie a LA TerreCâble secteur Tension d’utilisation secteurMontage Connexions OpérationTélédétection Connexion à la chargeConnexion en série ou en parallèle avec d’autres appareils Fusible NettoyageSicherheit WARNUNG! Dieses Gerät Muss Geerdet WerdenNetzkabel NetzbetriebsspannungAufstellung/Montage Anschlüsse BetriebVerbraucheranschluss LeistungsgrenzeIstwert-Fernerfassung Reihen- und Parallelschaltung mit anderen Geräten ÜberspannungsschutzSicherung WartungReinigung Sicurezza Attenzione Questo Strumento Deve Essere Collegato a TerraInstallazione Tensione d’esercizioCavo d’Alimentazione MontaggioCollegamento FunzionamentoLimite di potenza Allacciamento del caricoRilevamento a distanza Collegamento in serie/parallelo con altre unitàProtezione VentilazioneFusibile ManutenzionePulizia Seguridad ADVERTENCIA! Este Instrumento Debe Conectarse a TierraInstalación Voltaje de trabajo de alimentaciónCable de red MontajeConexiones FuncionamientoDetector remoto Conexión a la cargaConexión en serie o en paralelo a otras unidades Mantenimiento ProtecciónVentilación LimpiezaWarranty Information How do you get service?What does this warranty not cover? ProductReturn Material Authorization Policy Xantrex Technology Inc