Installation

Mains Operating Voltage

Check that the instrument operating voltage marked on the rear panel is suitable for the local supply. Should it be necessary to change the operating voltage, proceed as follows:

1.Ensure that the instrument is disconnected from the AC supply.

2.Remove the 6 screws holding the case upper and lift off the cover.

3.Change the plug-in connections between the transformer and main board following the diagram below:

230V OPERATION115V OPERATION

(VIEWFROMABOVE)

 

(VIEWFROMABOVE)

 

 

RED BROWN

 

BROWN

RED

 

 

L

BLUE

 

L

BLUE

P

 

P

 

N

 

N

 

(PARK)

 

(PARK)

 

 

 

 

 

4.Re-assemble in the reverse order.

5.To comply with safety standard requirements the operating voltage on the rear panel must be changed to clearly show the new voltage setting.

Mains Lead

When a three core mains lead with bare ends is provided this should be connected as follows:

BROWN

-

MAINS LIVE

 

BLUE

-

MAINS NEUTRAL

 

GREEN/YELLOW

-

EARTH

Safety Earth Symbol

When fitting a fused plug a 5 amp fuse should be fitted inside the plug. As the colours of the wires in the mains lead of this apparatus may not correspond with the coloured markings identifying the terminals in your plug proceed as follows:

The wire which is coloured green-and-yellow must be connected to the terminal in the plug which is marked by the letter E or by the safety earth symbol shown above or coloured green or green- and-yellow.

The wire which is coloured blue must be connected to the terminal which is marked with the letter N or coloured black.

The wire which is coloured brown must be connected to the terminal which is marked with the letter L or coloured red.

WARNING! THIS INSTRUMENT MUST BE EARTHED.

Any interruption of the mains earth conductor inside or outside the instrument will make the instrument dangerous. Intentional interruption is prohibited.

Mounting

This instrument is suitable both for bench use and rack mounting. It is delivered with feet for bench mounting. The front feet include a tilt mechanism for optimal panel angle.

7

Image 8
Contents XPF Series Dual Output 35V 10A Powerflex DC Power Supply XPFPart Number Contact InformationTable of Contents Specification Output SpecificationsMeter Specifications Safety Emissions ImmunityInstallation Mains Operating VoltageMains Lead MountingConnections Operation Connection to the Load Remote SensingSeries or Parallel connection with other units ProtectionMaintenance CleaningVentilation FuseEquipment Required Access to Calibration AdjustmentsCalibration Set Power Limit Voltage CalibrationCurrent Calibration OVP CalibrationSécurité AVERTISSEMENT! CET Instrument Doit Etre Relie a LA TerreMontage Tension d’utilisation secteurCâble secteur Connexions OpérationConnexion en série ou en parallèle avec d’autres appareils Connexion à la chargeTélédétection Fusible NettoyageSicherheit WARNUNG! Dieses Gerät Muss Geerdet WerdenAufstellung/Montage NetzbetriebsspannungNetzkabel Anschlüsse BetriebIstwert-Fernerfassung LeistungsgrenzeVerbraucheranschluss Reihen- und Parallelschaltung mit anderen Geräten ÜberspannungsschutzReinigung WartungSicherung Sicurezza Attenzione Questo Strumento Deve Essere Collegato a TerraInstallazione Tensione d’esercizioCavo d’Alimentazione MontaggioCollegamento FunzionamentoLimite di potenza Allacciamento del caricoRilevamento a distanza Collegamento in serie/parallelo con altre unitàProtezione VentilazionePulizia ManutenzioneFusibile Seguridad ADVERTENCIA! Este Instrumento Debe Conectarse a TierraInstalación Voltaje de trabajo de alimentaciónCable de red MontajeConexiones FuncionamientoConexión en serie o en paralelo a otras unidades Conexión a la cargaDetector remoto Mantenimiento ProtecciónVentilación LimpiezaWarranty Information How do you get service?What does this warranty not cover? ProductReturn Material Authorization Policy Xantrex Technology Inc