Xantrex Technology XPF 35-10 manual Emissions, Immunity

Page 7

EMC

This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC. Compliance was demonstrated by meeting the test limits of the following standards:

Emissions

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were:

a)Radiated: Class B

b)Conducted: Class B

c) Harmonics: EN61000-3-2 (2000) Class A; the instrument is Class A by product category.

Immunity

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use.

Test methods, limits and performance achieved were:

a)EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A.

b)EN61000-4-3 (1997) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance A.

c)EN61000-4-11 (1994) Voltage Interrupt, 1 cycle, 100%, Performance B.

d)EN61000-4-4 (1995) Fast Transient, 1kV peak (AC line), 0.5kV peak (DC Outputs), Performance A.

e)EN61000-4-5 (1995) Surge, 0.5kV (line to line), 1kV (line to ground), Performance A.

f)EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output connections <3m not tested), Performance A.

According to EN61326 the definitions of performance criteria are:

Performance criterion A: ‘During test normal performance within the specification limits.’

Performance criterion B: ‘During test, temporary degradation, or loss of function or performance which is self-recovering’.

Performance criterion C: ‘During test, temporary degradation, or loss of function or performance which requires operator intervention or system reset occurs.’

Where Performance B is stated it is because DC Output regulation may deviate beyond Specification limits under the test conditions. However, the possible deviations are still small and unlikely to be a problem in practice.

Note that if operation in a high RF field is unavoidable it is good practice to connect the PSU to the target system using screened leads which have been passed (together) through an absorbing ferrite sleeve fitted close to the PSU terminals.

Cautions

To ensure continued compliance with the EMC directive observe the following precautions:

a)after opening the case for any reason ensure that all signal and ground connections are remade correctly and that case screws are correctly refitted and tightened.

b)In the event of part replacement becoming necessary, only use components of an identical type, see the Service Manual.

6

Image 7
Contents XPF XPF Series Dual Output 35V 10A Powerflex DC Power SupplyContact Information Part NumberTable of Contents Output Specifications SpecificationMeter Specifications Safety Immunity EmissionsMounting InstallationMains Operating Voltage Mains LeadConnections Operation Protection Connection to the LoadRemote Sensing Series or Parallel connection with other unitsFuse MaintenanceCleaning VentilationCalibration Access to Calibration AdjustmentsEquipment Required OVP Calibration Set Power LimitVoltage Calibration Current CalibrationAVERTISSEMENT! CET Instrument Doit Etre Relie a LA Terre SécuritéCâble secteur Tension d’utilisation secteurMontage Opération ConnexionsTélédétection Connexion à la chargeConnexion en série ou en parallèle avec d’autres appareils Nettoyage FusibleWARNUNG! Dieses Gerät Muss Geerdet Werden SicherheitNetzkabel NetzbetriebsspannungAufstellung/Montage Betrieb AnschlüsseVerbraucheranschluss LeistungsgrenzeIstwert-Fernerfassung Überspannungsschutz Reihen- und Parallelschaltung mit anderen GerätenSicherung WartungReinigung Attenzione Questo Strumento Deve Essere Collegato a Terra SicurezzaMontaggio InstallazioneTensione d’esercizio Cavo d’AlimentazioneFunzionamento CollegamentoCollegamento in serie/parallelo con altre unità Limite di potenzaAllacciamento del carico Rilevamento a distanzaVentilazione ProtezioneFusibile ManutenzionePulizia ADVERTENCIA! Este Instrumento Debe Conectarse a Tierra SeguridadMontaje InstalaciónVoltaje de trabajo de alimentación Cable de redFuncionamiento ConexionesDetector remoto Conexión a la cargaConexión en serie o en paralelo a otras unidades Limpieza MantenimientoProtección VentilaciónHow do you get service? Warranty InformationProduct What does this warranty not cover?Return Material Authorization Policy Xantrex Technology Inc