Xantrex Technology RS232-XHR manual Lockout. See also Remote/Local Operation on

Page 48

Operation

Command Reference

Command

Description

FOLD?

Asks for the supply’s present foldback setting.

 

Response: FOLD <mode> where mode is:

 

0 (OFF) or

 

1 (CV or Constant Voltage mode) or

 

2 (CC or Constant Current mode)

GTL

Go to local. Returns the power supply to local control. See LLO (Local

 

Lockout). See also “Remote/Local Operation” on page 27.

HOLD <1/ON>,<0/OFF>

Enables or disables voltage/current setting hold mode for the supply. When

 

HOLD ON is specified, hold mode is enabled so that all voltage and current

 

settings which would normally be implemented by the supply are held until a

 

TRG (trigger) command is received. This feature allows you to synchronize

 

the operation of several supplies.

 

Initial value: HOLD OFF or HOLD 0

HOLD?

Asks for the present hold mode setting.

 

Response: HOLD 0 (OFF or disabled) or

 

HOLD 1 (ON or enabled)

ID?

Asks for the power supply model and the master EPROM version.

 

Response: ID <model name><version>

 

 

IDATA <Ilo>,<Ihi>

Calculates and records the slope and offset for programmed current using

 

ILO and IHI data. Set CMODE ON before using this command. See also the

 

calibration procedures in Section 4.

 

<Ilo> and <Ihi> are in <current> format.

IHI

In response to this command, the power supply sends a programmed

 

current value to the output terminal. This value is at the high end of the

 

power supply’s current range and is read by an external device connected

 

as part of the calibration procedure. Refer to this value as IHI and record it

 

to use as input with the IDATA command. Set CMODE ON before using this

 

command. See also the calibration procedures in Section 4.

ILO

In response to this command, the power supply sends a programmed

 

current value to the output terminal. This value is at the low end of the

 

power supply’s current range and is read by an external device connected

 

as part of the calibration procedure. Refer to this value as ILO and record it

 

to use as input with the IDATA command. Set CMODE ON before using this

 

command. See also the calibration procedures in Section 4 .

IMAX <current>

Sets an upper soft limit on the supply’s programmed output current. If the

 

soft limit is exceeded, or if the soft limit value is lower than the present

 

output current setting, the supply will ignore the command, turn on the

 

ERR LED, and set the ERR bit in the bit registers.

 

Range: 0 to model maximum output current (IMAX)

 

Initial value: model IMAX

IMAX?

Asks for the supply’s soft current limit setting.

 

Response: IMAX <current>

 

 

46

Operating Manual for RS-232 for XHR/XFR Series Power Supply

Image 48
Contents RS232-XHR RS232-XFR Page Operating Manual for How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Calibration Features and Specifications DescriptionFeatures and Functions FeaturesFeatures and Specifications Readback FunctionsSpecifications Models 140 12-100 20-60 35-35 40-30Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceFlow Switch Setting Condition CommunicationsBaud Rate XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J216 Connections RS-232 Connector PinSwitch Position Power on results RemoteStartup Condition Default Settings Model Example IsetExample Command RS-232 ConnectionLocal Lockout Power Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J7 ConnectorJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure VoltageVoltage Readback Calibration Procedure Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc