Calibration

Over Voltage Protection (OVP) Calibration

Over Voltage Protection (OVP) Calibration

We recommend that you perform OVP calibration every six months. Connecting a digital voltmeter as in “Voltage Calibration Setup” is optional.

1.Disconnect all loads from the power supply.

2.Ensure that jumper J65 on the interface PCB is CLOSED to enable remote OVP calibration (Jumper J65 is closed at the factory). See Figure 2.3 RS-232 Interface PCB, p. 22 for the location on the PCB of jumper J65.

3.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

4.Send the command OVCAL to the power supply. The ADDR LED will light during OVP calibration. Calibration is complete when the ADDR LED turns off. This may take a few minutes.

5.Use the OVSET, OVSET?, and VSET commands to trip the OVP level, confirming that the calibration was successful. When you trip the OVP level, the red OVP LED will light and the voltage will drop to zero. Send the command RST to clear the OVP condition. Refer to the OVP program accuracy specification in Section 1.

6.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

60

Operating Manual for RS-232 for XHR/XFR Series Power Supply

Image 62
Contents RS232-XHR RS232-XFR Page Operating Manual for How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Calibration Features and Specifications DescriptionFeatures and Specifications Features and FunctionsFeatures Readback FunctionsSpecifications Models 140 12-100 20-60 35-35 40-30Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceBaud Rate Flow Switch Setting ConditionCommunications XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J216 Connections RS-232 Connector PinStartup Switch Position Power on resultsRemote Condition Default Settings Model Example IsetExample Local Lockout CommandRS-232 Connection Power Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J7 ConnectorJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Program Calibration Procedure Voltage Mode CalibrationSetup VoltageVoltage Readback Calibration Procedure Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc