Xantrex Technology RS232-XHR manual Auxa?

Page 44

Operation

Command Summary

Table 3.4

Query Commands

Command

Description

 

 

AUXA?

Asks for the state of the set value for the AUXA command

 

 

AUXB?

Asks for the state of the set value for the AUXB command

 

 

CMODE?

Asks for the power supply’s calibration mode status.

 

 

DLY?

Asks for the programmable time delay setting before the supply

 

reports fault conditions.

 

 

ERR?

Asks for the most recent remote programming error which occurred in

 

the supply since the last time the error query command (ERR?) was

 

used.

 

 

FOLD?

Asks for the supply’s present foldback setting.

 

 

HOLD?

Asks for the present hold mode setting.

 

 

ID?

Asks for the power supply’s model name and master EPROM version.

 

 

IMAX?

Asks for the supply’s soft current limit setting.

 

 

IOUT?

Measures the supply’s actual current output.

 

 

ISET?

Asks for the supply’s present output current limit setting.

 

 

OUT?

Asks for the present enabled/disabled status of the supply’s output.

 

 

OVSET?

Asks for the supply’s present over voltage protection limit.

 

 

REN?

Asks for the state of remote enable.

 

 

ROM?

Asks for the version number of the master and slave EPROMs on the

 

interface PCB.

VMAX?

Asks for the supply’s soft voltage limit setting.

 

 

VOUT?

Measures the supply’s actual voltage output.

 

 

VSET?

Asks for the supply’s present output voltage setting.

 

 

42

Operating Manual for RS-232 for XHR/XFR Series Power Supply

Image 44
Contents RS232-XHR RS232-XFR Page Operating Manual for How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Calibration Features and Specifications DescriptionFeatures and Functions FeaturesFeatures and Specifications Readback FunctionsSpecifications Models 140 12-100 20-60 35-35 40-30Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceBaud Rate Flow Switch Setting ConditionCommunications XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J216 Connections RS-232 Connector PinStartup Switch Position Power on resultsRemote Condition Default Settings Model Example IsetExample Local Lockout CommandRS-232 Connection Power Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J7 ConnectorJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure VoltageVoltage Readback Calibration Procedure Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc