Command

Description

VLO

In response to this command, the power supply sends a programmed

 

voltage value to the output terminal. This value is at the low end of the

 

power supply’s voltage range and is read by an external voltmeter

 

connected as part of the calibration procedure. Refer to this value as VLO

 

and record it to use as input with the VDATA command. Set CMODE ON

 

before using this command. See also the calibration procedures in

 

Section 4.

VMAX <voltage>

Sets an upper soft limit on the supply’s programmed output voltage. If the

 

soft limit is exceeded, or if the soft limit value is lower than the present

 

output voltage setting, the supply will ignore the command, turn on the

 

ERR LED, and set the ERR bit in the accumulated status register.

 

Range: 0 to model maximum output voltage (VMAX)

 

Initial value: model VMAX

VMAX?

Asks for the supply’s soft voltage limit setting.

 

Response: VMAX <voltage>

 

 

VOUT?

Measures the supply’s actual voltage output using the built-in voltage

 

readback circuitry.

 

Response: VOUT <voltage>

 

 

VRDAT <Vlo>,<Vhi>

Calculates and records the slope and offset for readback voltage using

 

VRLO and VRHI data. Set CMODE ON before using this command. See

 

also the calibration procedures in Section 4 .

 

<Vlo> and <Vhi> are in <voltage> format.

VRHI

The power supply outputs a voltage value to an external voltmeter

 

connected as part of the calibration procedure and records a voltage

 

readback value internally. These values are at the high end of the

 

programmed voltage range. Refer to the output value as VRHI and record it

 

to use as input with the VRDAT command. Set CMODE ON before using

 

this command. See also the calibration procedures in Section 4.

VRLO

The power supply outputs a voltage value to an external voltmeter

 

connected as part of the calibration procedure and records a voltage

 

readback value internally. These values are at the low end of the

 

programmed voltage range. Refer to the output value as VRLO and record

 

it to use as input with the VRDAT command. Set CMODE ON before using

 

this command. See also the calibration procedures in Section 4.

VSET <voltage>

Sets the power supply’s output voltage in volts (default) or in millivolts. This

or

programmed voltage is the actual output in CV (constant voltage) mode or

VSET <-voltage>

the voltage limit in CC (constant current) mode.

 

If you enter a negative voltage value, the power supply will assert a signal

 

on the J7-4 user signal line. You can use the user signal to trip external

 

relays to switch the output polarity.

 

Range: 0 to model maximum output voltage (VMAX)

 

Initial value: 0 volts

 

 

VSET?

Asks for the power supply’s present output voltage setting. Does not apply

 

to voltage settings which are being held. See HOLD command.

 

Response: VSET <voltage>

 

 

50

Operating Manual for RS-232 for XHR/XFR Series Power Supply

Image 52
Contents RS232-XHR RS232-XFR Page Operating Manual for How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Calibration Features and Specifications DescriptionFeatures and Functions FeaturesFeatures and Specifications Readback FunctionsSpecifications Models 140 12-100 20-60 35-35 40-30Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceCommunications Flow Switch Setting ConditionBaud Rate XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J216 Connections RS-232 Connector PinRemote Switch Position Power on resultsStartup Condition Default Settings Model Example IsetExample RS-232 Connection CommandLocal Lockout Power Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J7 ConnectorJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure VoltageVoltage Readback Calibration Procedure Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc