Xantrex Technology RS232-XHR manual XON Switch Setting Condition

Page 27

Installation and Configuration

Communications

Once flow control is selected as option, two flow control methods are available. You can use the rear panel XON switch to select either software-based XON/XOFF flow protocol or hardware-based CTS/RTS or DTR/DSR flow protocol. Table 2.4 shows the flow control conditions with regards to the markings on the rear panel subplate. For a diagram of the rear panel subplate, see Figure 2.2, on page 21.

Table 2.4 Hardware/Software Flow Control Selection

XON Switch Setting

Condition

1

Hardware Flow Control

 

(CTS/RTS or DTR/DSR protocol)

0

Software Flow Control

 

(XON/XOFF protocol)

Software This software protocol uses special characters that synchronize device

Flow Control communications. The device sends the control code XOFF (ASCII value 13 hex) when it is not ready to receive data. The device sends the control code XON (ASCII value 11) when it is ready to resume receiving data.

To choose XON/XOFF protocol, select the XON switch position shown in Table 2.4.

Hardware Hardware flow control protocol requires a dedicated control line on the RS-232 Flow Control cable. The dedicated control line is needed so that the device can signal the controller

when it is ready to receive data. You can designate either RTS/CTS or DTR/DSR protocol for the RS-232 interface.

Set the XON switch for hardware control using the information in Table 2.4. With hardware flow control selected, RS-232 Interface PCB jumpers J217 and J218 dictate the type of flow protocol. Table 2.5, “Jumper/Pin Settings for RTS/CTS Flow Control (Default)” and Table 2.6, “Jumper/Pin Settings for DTR/DSR Flow Control (Default)” identify jumper connections and input and output pin designations for each protocol. See also “Transmit / Receive Pin Selection” on page 26 for RS-232 connector pinouts.

Note:When resetting the flow control switches during a working session, turn the power supply OFF and then ON again to ensure that the new settings are put into effect.

Table 2.5 Jumper/Pin Settings for RTS/CTS Flow Control (Default)

 

Jumper Connections

RS-232 Connector Pin

RTS (Input to RS-232)

J218 to 2-4 and 3-5

7

 

 

 

CTS (Output to RS-232)

J217 to 1-3 and 4-6

8

 

 

 

Release 3.1

25

Image 27
Contents RS232-XHR RS232-XFR Page Operating Manual for Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyManual Revisions Main SectionsAbout This Manual Contents Calibration Description Features and SpecificationsReadback Functions Features and FunctionsFeatures Features and SpecificationsModels 140 12-100 20-60 35-35 40-30 SpecificationsModels 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Initial Inspection IntroductionInitial Inspection Installation and ConfigurationRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Step # Description Action Reference Basic Setup ProcedureFlow Switch Setting Condition CommunicationsBaud Rate Jumper Connections RS-232 Connector Pin XON Switch Setting ConditionJumper J216 Connections RS-232 Connector Pin Power Supply DCE Host Computer DTE Pin connectorSwitch Position Power on results RemoteStartup Iset Condition Default Settings Model ExampleExample Command RS-232 ConnectionLocal Lockout Parameter Setting Power Supply Settings12 OVP Control Mode Selection User Options and SettingsHigh OFF LOW User Signals J7 Connector User SignalsJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block RS-232 Operation Command SyntaxFormat OperationFloating Number Definition Example Parameter Description FormOperation Command Description Command SummaryAUXA? Cmode Auxa 1 on Command ReferenceResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers 4096 Condition Mnemonic Bit Position Bit WeightError # Error Identification Explanation Error CodesLEDs TroubleshootingCalibration Voltage Voltage Mode CalibrationSetup Voltage Program Calibration ProcedureVoltage Readback Calibration Procedure Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc