Xantrex Technology RS232-XHR manual Status, and Fault Registers on

Page 51

 

Operation

 

Command Reference

 

 

 

Command

Description

 

STS?

Asks for the supply’s present status register. Status conditions are stored in

 

the status register. Each bit represents a separate condition. When the

 

condition is true, the corresponding bit is 1 (true). Bits remain set in the

 

status register as long as the condition is true. See “Accumulated Status,

 

Status, and Fault Registers” on page 51.

 

Response: STS <status mask> where status mask is the decimal equivalent

 

of the total bit weights for the operating conditions as listed in the status

 

register.

 

TRG

Implements programmed voltage and current settings which had been in

 

hold mode. The supply operates with previous values until the TRG (trigger)

 

command is sent.

 

UNMASK <mnemonics>

Enables you to select the supply operating conditions that you are most

 

interested in monitoring for fault occurrence. Mnemonics describing the

 

conditions are separated from each other by commas, and may be sent in

 

any order.

 

Specifying one or more mnemonics which describe the conditions (or the

 

decimal equivalent of their total bit weight) enables the selected conditions

 

to set bits in the supply’s fault and status registers during operation. A bit is

 

set in the fault register when the corresponding bit in the status register

 

changes from 0 to 1 and the corresponding bit in the mask register is 1. See

 

“Accumulated Status, Status, and Fault Registers” on page 51.

 

Mnemonics: CV, CC, OV, OT, SD, FOLD, ERR, PON, REM, ACF, OPF,

 

SNSP, ALL, NONE

 

Initial value: UNMASK NONE

 

 

 

UNMASK?

Asks for the supply's fault conditions which are currently enabled

 

(unmasked).

 

Response: UNMASK <fault mask> where fault mask is the decimal

 

equivalent of the total bit weights for the operating conditions as listed in the

 

status and fault registers See “Accumulated Status, Status, and Fault

 

Registers” on page 51.

 

VDATA <Vlo>,<Vhi>

Calculates and records the slope and offset for programmed voltage using

 

VLO and VHI data. Set CMODE ON before using this command. See also

 

the calibration procedures in Section 4.

 

<Vlo> and <Vhi> are in <voltage> format.

 

VHI

In response to this command, the power supply sends a programmed

 

voltage value to the output terminal. This value is at the high end of the

 

power supply’s voltage range and is read by an external device connected

 

as part of the calibration procedure. Refer to this value as VHI and record it

 

to use as input with the VDATA command. Set CMODE ON before using

 

this command. See also the calibration procedures in Section 4.

 

Release 3.1

49

Image 51
Contents RS232-XHR RS232-XFR Page Operating Manual for Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyManual Revisions Main SectionsAbout This Manual Contents Calibration Description Features and SpecificationsReadback Functions Features and FunctionsFeatures Features and SpecificationsModels 140 12-100 20-60 35-35 40-30 SpecificationsModels 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Initial Inspection IntroductionInitial Inspection Installation and ConfigurationRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Step # Description Action Reference Basic Setup ProcedureFlow Switch Setting Condition CommunicationsBaud Rate Jumper Connections RS-232 Connector Pin XON Switch Setting ConditionJumper J216 Connections RS-232 Connector Pin Power Supply DCE Host Computer DTE Pin connectorSwitch Position Power on results RemoteStartup Iset Condition Default Settings Model ExampleExample Command RS-232 ConnectionLocal Lockout Parameter Setting Power Supply Settings12 OVP Control Mode Selection User Options and SettingsHigh OFF LOW User Signals J7 Connector User SignalsJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block RS-232 Operation Command SyntaxFormat OperationFloating Number Definition Example Parameter Description FormOperation Command Description Command SummaryAUXA? Cmode Auxa 1 on Command ReferenceResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers 4096 Condition Mnemonic Bit Position Bit WeightError # Error Identification Explanation Error CodesLEDs TroubleshootingCalibration Voltage Voltage Mode CalibrationSetup Voltage Program Calibration ProcedureVoltage Readback Calibration Procedure Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc