Xantrex Technology RS232-XHR manual Voltage Mode Calibration, Setup

Page 58

Calibration

Voltage Mode Calibration

Voltage Mode Calibration

Voltage

1.

Disconnect the load from the power supply which is to be calibrated.

Calibration

2.

Connect a voltmeter across the power supply’s output terminals.

Setup

 

 

 

 

 

 

 

 

 

 

 

 

Power Supply

Power Supply

 

 

 

Positive Output

Negative Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Voltage

Program

Calibration

Procedure

+V -

VOLTMETER

Figure 4.1 Voltage Calibration Setup

1.Set the power supply for calibration as in Figure 4.1.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command VLO; ILO to the power supply. Measure and record the output shown on the external voltmeter.

4.Send command VHI; IHI to the supply. Measure and record the output voltage as shown on the external voltmeter.

5.Send the command VDATA <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter when the VLO and VHI commands were sent. When the power supply is calibrated, the low to high voltage program calibration values are stored as constants.

6.Program the supply at various levels using the VSET command to confirm that the calibration was successful and that linearity is observed. See the voltage program accuracy specification in Section 1.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

56

Operating Manual for RS-232 for XHR/XFR Series Power Supply

Image 58
Contents RS232-XHR RS232-XFR Page Operating Manual for How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Calibration Features and Specifications DescriptionFeatures and Specifications Features and FunctionsFeatures Readback FunctionsSpecifications Models 140 12-100 20-60 35-35 40-30Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceCommunications Flow Switch Setting ConditionBaud Rate XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J216 Connections RS-232 Connector PinRemote Switch Position Power on resultsStartup Condition Default Settings Model Example IsetExample RS-232 Connection CommandLocal Lockout Power Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J7 ConnectorJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Program Calibration Procedure Voltage Mode CalibrationSetup VoltageVoltage Readback Calibration Procedure Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc