Omega Speaker Systems RD8300 Modbus Settings, Mouse Pointer Enable, Config Auto Save, Unit Tag

Page 82

Section 5 – Recorder Setup

5.7.2.2Modbus Settings

This is only applicable if the serial port option is installed in the unit. Modbus is a communication protocol which runs over RS232 (single unit) or RS485 (multiple units). The ID uniquely identifies the device in a multidrop environment and can be set from 1 to 255. Note that this does not imply that 255 units can be on line at a time. The limit is usually 32 to 64. The Baud Rate is the serial communication speed which must match that of the host controller. This can be set from 9600 to 115200 baud.

5.7.2.3Mouse Pointer Enable

Check this Enable checkbox if you plan on using an external mouse or other pointing device. This will keep the mouse pointer or cursor visible at all times. Note: This setting only takes effect AFTER you power down the unit and restart it.

5.7.2.4Config Auto Save

Check this Enable checkbox to have the Configuration file saved automatically each time a change is made to the unit configuration. If you are setting the system up for the first time you can disable this feature by unchecking this box so the process is a little quicker, however DO NOT FORGET TO SAVE THE CONFIGURATION using the Quick Save button in the Utility Setup Menu.

5.7.2.5Menu Timeout

The Menu Timeout sets the time in seconds after which any menu visible on the screen will collapse and the display will revert back to the live display. Select 60 to 300 seconds from the dropdown box.

5.7.2.650/60 Hz Filter

In noisy environments the AC mains frequency interference can affect the readings on the most sensitive (millivolt) ranges when running at sample rates asynchronous to the AC mains frequency. If the sample rate can be made to run synchronously with the AC mains frequency, the noise effect of that frequency is minimized because the recorder has differential inputs and samples at the same place on AC wave for every sample. This does however place a limit on what the sample rates can be. The common denominator for 50 and 60 Hz is a sample rate of 16.66 samples per second to eliminate the AC hum. This means that the fastest sample rate to meet this requirement is around 1.5 to 2.2 samples per second depending on the number of analog input channels in the recorder. Checking the box will activate the filter but will limit the maximum sample rate irrespective of what is selected elsewhere. This is usually not an issue for thermocouples and RTDs which are typically the most affected by AC noise.

5.7.2.7Unit Tag

The Unit Tag identifies the recorder. It appears in the task bar and all data files. Enter up to 22 characters. The unit tag is Process Recorder in the figure above.

Once the changes have been made press the accept button to exit or the cancel button to quit without saving the changes.

5.7.3Date/Time Setting

Tapping the Date/Time icon shown left will bring up the Date/Time dialog. Here the user can set the current Date, Time and Time Zone as well as

the date format

To change the time – tap on the hour, min, secs accordingly then adjust using the up and down arrow keys. Select the current time zone from the pull down menu

Fig 5.7.3 – Date – Time Setup

Page 73

09/11/2008 REV 0.99

Image 82
Contents User’s Guide Servicing North America Table of Contents Wiring Specifications and Procedures Environment and Location InstallationDisplay Elements Browsing Historic Data Timer/Totalizer Reset Special EventsTurning the unit off Annotating Trends Browsing AnnotationsSystem Administrative Menu AlarmsRecord Setup Digital Input/OutputUSB Connections Utility Programs Touch Screen Cleaning Instructions CalibrationWorld Wide Web Web Based FTP Manager Ethernet Protocols 102Symbols Trademarks09/11/2008 REV Overview General DescriptionSystem Overview System Overview09/11/2008 REV Recorder Front Display AreaIR Comm Port Status IndicatorMedia Door StylusRecorder Rear Panel Power SupplyAnalog Inputs Input/Output Module Computer InterfaceConfiguration Files FilesActivity Files Exhibitor SoftwareData Files Miscellaneous FilesQuick Start Guide Activity Lights KeyboardingHelp SecurityRemote Access Passwords User LevelsEquipment Handling Environment and LocationInstallation Recorder Dimensions in Inches mmPanel Mounting Front Panel Cut OutPanel Front Gasket Rear View Panel InsertionPower Requirements Wiring Specifications and ProceduresPower Connections Signal Input Wiring Rear PanelThermocouple Inputs Resistance Temperature Detector RTD InputsAnalog Inputs Linear InputsPotential Free Contacts FrequencyEthernet Relay Output, Contact InputAudio Connection USB ConnectionMouse and Keyboard Connection Isolated Serial Interface Option 24Vdc Isolated Output Option18 RS485 Adapter 19 RS485 120 Ohm Jumper Live Display Area Display ElementsTask Bar Utility Setup Menu Utility Setup Menu09/11/2008 REV Record Session Alarm, Event and Activity LogsFolder Mydata Data File Mydata.dat Bitmap Files Text Files Configuration Manager Load ConfigurationSending Email E-mail WindowDisplay Utilities User LoginDisplay Brightness Special Events Timer/Totalizer ResetRecord on Alarm Background ColorSystem Setup Email on Alarm EventRecord Setup Digital Input Setup Record Mode Alarm Master SetupAnnotating Trends Drawing on the ScreenBrowsing Historic Data Adding a Text MessageZoom and Compress Data 13.4 Zoom with point markersView Memos and Bitmaps Using the Cursor13.8 Cursor Mode 13.6 Memo View 13.7 Bitmap ViewUsing Browser Jump To Browsing AnnotationsSetup display as shown right Master Channel SetupInput Channels Channel Setup Channel SetupLinear Input Channel Type SetupResistance Input Industrial Square Root Extraction Thermocouples and RTDsCalculated Channel Setup Calculated Input09/11/2008 REV Gated Timer Conditional InputReset Point TotalizerFrequency/Pulse Count Frequency13 Pulse Count Setup Pulse CountRecorder Setup Display Builder Building the DisplayEditing 3 Display BuilderRotate List Digital Displays Digital Display SetupWidgets Will show OFFDisplaying Min, Max or Average Values Channel Type Shows the current widget style for Bar Graph SetupBar Graphs Channel. For information only09/11/2008 REV 09/11/2008 REV Trend Setup Trend Setup Window09/11/2008 REV Alarms RateAlarm Master Setup 1 Alarm Master SetupAlarm Setup Digital Input/Output Digital Output Setup 1 Digital Output SetupDigital Input Setup 09/11/2008 REV Record Setup Select ChannelsWhat to Record Media and FilenameRecord Rate Point type Max Min AverageSplit Files Timed RecordTriggered Record Media Alarm System Administrative Menu System Admin Menu09/11/2008 REV System Setup Temperature CompensationNTP Time Server Config Auto Save Modbus SettingsMouse Pointer Enable Menu TimeoutUser Setup Password Management UserFile Explorer Web User Manager Network Settings7 Web User Admin 6 Network IP AddressSet ActiveSync Password Smtp Email Settings9 USB ActiveSync Password Registry Languages System ReFlash09/11/2008 REV Communication World Wide Web3 View Recorder Web Based FTP Manager 2 FTP File Manager Communication Ethernet Protocols OPC ServerModbus Status Registers Modbus ServerModbus Data Registers Register # USB Connections Utility Programs FTP Manager09/11/2008 REV 09/11/2008 REV Remote Display Remote Control3 Remote Recorder 4 Remote Manager Remote Manager5 Remote Manager Connected Touch Screen Cleaning InstructionsCalibration Zero Cal Calibrating Linear Voltage InputsSet Channel Full CalCalibrating the RTD Current Source Store CalCalibrating Frequency Go to Table of Contents Power Requirements Resolution ColorVirtual Chart Speed Power Fail ProtectionType Range C Accuracy C Range F Accuracy F Weight SafetyDimensions Recording RatesAppendix B Digital Input Voltage Adjustment Input Board Fig B-4 Digital IO Module Jumpers 2 Channel shownWARRANTY/DISCLAIMER Temperature