SSI America 8040, 8020 manual Modify Input Offsets, Select Survey Template

Page 22

9. Modify Input Offsets

The Modify Input Offsets screen allows the user to modify the offsets of each channel, or to clear all of the offsets. To select a channel, highlight the “Select Input” line and enter a number between 1 and 40. To enter the offset value, highlight the “Enter offset value” line and enter a number between –1000 and 1000. The user can toggle the sign (+/-) of the number being entered by pressing the circular arrow button. To clear out all of the offsets, highlight the “Enter 1 here to clear all offsets” line and enter a 1.

10. Select Survey Template

The Select Survey Template screen allows the user to manually select a template to run a survey from. The SDS is capable of saving thirty-two templates per template file. Each template will be listed by template name. To create a template, go to the Template Manager screen section, Manage Survey Templates, using the SDS software and follow the procedures. To select a template, use the up or down arrow keys to highlight the template’s name and press Enter. The Display Survey Information screen will be displayed (menu option 2).

Super Systems Inc.

Page 22

SDS Data Logger Manual #4565 Rev D

Image 22
Contents SDS Data Logger Table of Contents Product Overview5SDS Software Appendix a Standard Calibration Range Revision History98 Product Specifications Product DescriptionSoftware Installation Calibration20 mA. Current Loop connections Thermocouple connectionsVoltage connections Getting StartedSurvey Templates Page Installation Installing the SDS SoftwareConnecting The SDS Data Logger To a Network Repair/Remove SDS Reporter screenIP Address Subnet Mask Default Gateway SDS 8020/8040 Data Logger Menu ListConfiguration Level Level 2 Pass Code Required Page Keypad Assignments Overview Display Display Survey InformationDisplay Input Settings Display Input Offsets Display Logged DataHelp Select Input ChannelsModify Input Settings NNMModify Input Offsets Select Survey TemplateBattery Status About/Sign-On Revision DisplaySet Display Values Language/LenguaQuick Chart Overview Display SelectedQuick Chart Setup Modify Survey InformationSet the Date and Time Port Setup Input CalibrationTC Type mV Range Chart Cold Junction Trim Firmware revision 1.09 or below Firmware Revision 1.10 and above Firmware Revision 1.11 and above To Perform a Cold Junction CalibrationFirmware Version 1.09 and below System SetupInput Board Version 1.08 and above Set Pass Codes Set IP AddressSurvey Deviation Alarm Log Clear Logged DataPage SDS Reporter File Æ New / New ButtonFile Menu Options File Æ Open / Open Button Report Properties TabPage Manually Creating a Survey Record Page 10 Report Properties Thermocouples Tab Page 14a Report Properties Furnace Tab Version 1.101.0.76 & below Page Page Page Page Page File Æ Export File Æ Save / File Æ Save As / Save ButtonFile Æ Print Page File Æ Exit Print ButtonFile Æ Print Preview SDS Æ Manage Survey TemplatesSDS Æ Download Surveys and Data Page Page SDS Æ Launch Direct FTP 36a FTP session window with Internet Explorer FTP Login For Access to the a Drive Ftp//LANGSSI772@IP AddressSDS Æ View Real-time Data Page SDS Æ Manage Survey Records 43 Confirm delete of local records SDS Æ Perform Comms Test 45 Changes Saved message boxPage Help Æ Check for Updates Options Æ Multiple SDS ModeOptions Æ Show Labels On Chart Options Menu Options54 No updates message box Help Æ About55 SDSReporter About box SDS Data TagsTC Survey Summary Tag Description UsageSdssp Sample Output Tag Tag Sample OutputTC Offsets Description Usage Tag Min/Max/Mean/Spread/Deviation Tag Description UsageCreating a Personalized Template Page Page SDS Template Manager New Button / File Î NewOpen Button / File Î Open Save Buttons / File Î Save File Î Save AsExit Button / File Î Exit Add New Template ButtonSave Changes to Template Button Delete Template Button Description of Template Manager Screen Tabs Template Information TabSurvey Information Tab Active TCs Tab Controller Information Tab SDS Reporter Versions 1.101.0.76 and above 18 New SDS Template ManagerSave Buttons / File Î Save Copy Button Add Template ButtonDelete Button Arrow ButtonsTemplate Information Tab Survey Information Tab Active TCs Tab Controller Information Tab Description of Template Manager Screen Fields Version 1.101.0.71 & Below General Notes Appendix a Standard Calibration Range Thermocouple Type Range in FRevision History