SSI America 8020, 8040 manual Cold Junction Trim

Page 31

be calibrated. To select the TCs to calibrate, enter a number between 0 and 31. To determine the number, follow these steps:

TC1 has a “value” of 1, TC2 has a “value” of 2, TC3 has a “value” of 4, TC4 has a “value” of 8, and TC5 has a “value” of 16.

A TC that is on has a value of 1. A TC that is off has a value of 0.

Multiply the TC’s value by its “value” and add all 5 TC values up.

For example, we wish to calibrate TC1, TC4, and TC5. TC1 = 1 (1 * 1), TC2 = 0 (0 * 2), TC3 = 0 (0 * 4), TC4 = 8 (1 * 8), and TC5 = 16 (1 * 16). The number to enter would be 1 + 0 + 0 + 8 + 16 = 25. The line will now look like this: CH’s to cal 1Y 2N 3N 4Y 5Y.

To select the range to calibrate, select one of the codes above (0 through 5) and press Enter. The mV ranges displayed above are ideal because they cover the mV range for the selected thermocouple type.

On the next line, enter a 0 for ZERO or a 1 for SPAN. Enter the target value (in MV) as a value ranging from 0 to 320. The SPAN value should be 90% of full scale. That means for 80mV, the span value should be 72mV. Scroll down to the next line and enter 1 to start the calibration. DO NOT START THE CALIBRATION until you have entered the proper value in the calibration device. Also be sure that the polarity of the wires is hooked up properly.

When the calibration is finished, the “Enter 1 here to start” line will read “Ready 0”. Pressing ESC will display the Input Calibration first screen (menu option 22).

The calibration process can be run multiple times and if initial calibration is off significantly, running multiple times is recommended.

23. Cold Junction Trim

The instructions for performing the Cold Junction Trim on the inputs will depend on the firmware level of the instrument that you have. You can determine the firmware version be checking menu option 13 Revision Display. The number to the right of the first line, “Main Processor”, will be the firmware level. Please select the instructions below based on the firmware level of the instrument.

Super Systems Inc.

Page 31

SDS Data Logger Manual #4565 Rev D

Image 31
Contents SDS Data Logger Product Overview5 Table of ContentsSDS Software Appendix a Standard Calibration Range Revision History98 Product Description Product SpecificationsCalibration Software InstallationGetting Started Thermocouple connectionsVoltage connections 20 mA. Current Loop connectionsSurvey Templates Page Installing the SDS Software InstallationRepair/Remove SDS Reporter screen Connecting The SDS Data Logger To a NetworkIP Address Subnet Mask Default Gateway Menu List SDS 8020/8040 Data LoggerConfiguration Level Level 2 Pass Code Required Page Keypad Assignments Display Survey Information Overview DisplayDisplay Input Settings Display Logged Data Display Input OffsetsSelect Input Channels HelpNNM Modify Input SettingsSelect Survey Template Modify Input OffsetsBattery Status Revision Display About/Sign-OnLanguage/Lengua Set Display ValuesOverview Display Selected Quick ChartModify Survey Information Quick Chart SetupSet the Date and Time Input Calibration Port SetupTC Type mV Range Chart Cold Junction Trim Firmware revision 1.09 or below Firmware Revision 1.10 and above To Perform a Cold Junction Calibration Firmware Revision 1.11 and aboveFirmware Version 1.09 and below System SetupInput Board Version 1.08 and above Set IP Address Set Pass CodesClear Logged Data Survey Deviation Alarm LogPage SDS Reporter File Æ New / New ButtonFile Menu Options Report Properties Tab File Æ Open / Open ButtonPage Manually Creating a Survey Record Page 10 Report Properties Thermocouples Tab Page 14a Report Properties Furnace Tab Version 1.101.0.76 & below Page Page Page Page Page File Æ Export File Æ Save / File Æ Save As / Save ButtonFile Æ Print Page SDS Æ Manage Survey Templates Print ButtonFile Æ Print Preview File Æ ExitSDS Æ Download Surveys and Data Page Page SDS Æ Launch Direct FTP 36a FTP session window with Internet Explorer Ftp//LANGSSI772@IP Address FTP Login For Access to the a DriveSDS Æ View Real-time Data Page SDS Æ Manage Survey Records 43 Confirm delete of local records 45 Changes Saved message box SDS Æ Perform Comms TestPage Options Menu Options Options Æ Multiple SDS ModeOptions Æ Show Labels On Chart Help Æ Check for UpdatesHelp Æ About 54 No updates message boxSDS Data Tags 55 SDSReporter About boxTag Description Usage TC Survey SummarySdssp Tag Sample Output Sample Output TagMin/Max/Mean/Spread/Deviation Tag Description Usage TC Offsets Description Usage TagCreating a Personalized Template Page Page New Button / File Î New SDS Template ManagerOpen Button / File Î Open File Î Save As Save Buttons / File Î SaveAdd New Template Button Exit Button / File Î ExitSave Changes to Template Button Delete Template Button Template Information Tab Description of Template Manager Screen TabsSurvey Information Tab Active TCs Tab Controller Information Tab 18 New SDS Template Manager SDS Reporter Versions 1.101.0.76 and aboveSave Buttons / File Î Save Arrow Buttons Add Template ButtonDelete Button Copy ButtonTemplate Information Tab Survey Information Tab Active TCs Tab Controller Information Tab Description of Template Manager Screen Fields Version 1.101.0.71 & Below General Notes Thermocouple Type Range in F Appendix a Standard Calibration RangeRevision History