Follow these steps to adjust the Z-arm to the proper height.
1Turn off power to the liquid handler.
2Locate the appropriate Z-height adjustment tool in the accessory package. Two Z-height adjustment tools are supplied in the package.
Part number Description
25051094 125 mm Z-height adjustment tool. Use this tool to adjust the Z-arm on the Quad-Z 215 Liquid Handler for liquid handling and injection.
25051095 175 mm Z-height adjustment tool. Use this tool to adjust the Z-arm on the Quad-Z 215 Liquid Handler for liquid handling.
3Place the Z-height adjustment tool on one of its ends near the center of the locator plate (if necessary, remove any racks or accessories before doing this).