Installation
Electrical Connections
2
Baud rate/mode
As a default, the baud rate/mode is set to 6, indicating that the liquid handler is set for a baud rate of 19200 and is a master device.
You’ll need to change the selection to 0 (zero) if the liquid handler is connected via the GSIOC to a Gilson system and is being controlled by the 506C System Interface from Gilson control software. A setting of 0 indicates the liquid handler is a slave device and the baud rate is being clocked externally.
Note that other selections are available if the liquid handler is being controlled by
To change the baud rate/mode:
1Gently insert a small flat blade screwdriver into the SW2 selector on the rear panel and turn it.
2Align the white dot with one of the indicated numbers.