Gilson 215 manual Installing different type of probe

Page 66

Maintenance

Replacing Parts

4

Installing different type of probe

To install a replacement probe of a different type than is currently installed, you may want to obtain a new probe guide insert for precise XY probe accuracy.

1Remove the transfer tubing’s 1/4"-28 fitting connected to the top of the isolation probe holder.

2Grasp the current probe and push it up through the top of the isolation probe holder.

3Remove the current probe guide insert by turning it counterclockwise. Then place the new probe guide insert into probe guide and secure it by turning it clockwise.

4Insert the new probe into the top of the isolation probe holder and pull it through the holder and the new probe guide insert until the tip of the probe is in the probe guide.

5Replace and tighten the 1/4"-28 fitting.

4-10

Image 66
Contents Quad-Z 215 Liquid Handler Users Guide Page Table of Contents Maintenance OperationAppendix C Gsioc Commands TroubleshootingAppendix a Replacement Parts and Accessories Appendix B RacksPage Declaration of Conformity Page Introduction1 Description Unpacking Standard Equipment Customer Service Technical Specifications Introduction Introduction Installation2 Locator Plate Installation InstallationArmlock Removal Arm Installation Installation Installing the Z-Arm Cable Support Rod Probe Installation Installing the Probe Guide InsertsInstalling the Probes Transfer Tubing Installation Plumbing ConnectionsInstalling the Level Sensing Cables To install the level sensing cablePlumbing Connections Installing the rinse station Rack Setup Code 200-Series Code 500-Series RacksCode 200-Series and Code 20-Series or Code 30-Series Racks Installation Only Code 20-Series or Code 30-Series Racks Final Z-Arm Adjustment Part number DescriptionFinal Z-ArmAdjustment Electrical Connections Rear PanelContact outputs Installation Input/Output PortsContact inputs Items you’ll needMaking connections Installation RS-232 Port Connecting an RS-232 cableGsioc Port Gsioc cableInstallation Unit ID and Baud Rate/Mode Selection Unit IDBaud rate/mode Installation Fuses Power Cord Connection Operation Front Panel Start Up Running Programs Configuring the Liquid Handler About Operation Pump Type Syringe OptionsRinse Site Operation Liquid Detector Adjust XY Model 841 X-coordinate 475.3 mm Y-coordinate 3.8 mm Height Operation Home Phase Safety Contact Bearing LifeOperation Sound and Display Emergency ContactTesting the Liquid Handler’s Contacts Maintenance4 Maintenance Helpful HintsCleaning Cleaning the Liquid HandlerCleaning the Fluid Path Cleaning methods Maintenance Maintenance Horizontal Pitch and Vertical Rods ProceduresLubricating What you needMm spacing Replacing Tubing Replacing PartsInstalling same type of probe Replacing a ProbeInstalling different type of probe Maintenance Replacing a Level Sensing Cable Removing the level sensing cableInstalling the new level sensing cable Maintenance Replacing a Fuse Checking Position Alignment Transporting the Liquid Handler Page Troubleshooting5 Error Messages TroubleshootingPosition sensor inactive Mechanical Probes no longer finding tube centerElectrical Input functions not operatingOutput functions not operating Unit not operational Unit blows fusesLiquid Level Detector Liquid level detector not detecting liquid levelLiquid level detector falsely detecting liquid level Repair and Return Policies Warranty repairNon-warranty repair Before calling usReturn procedure Rebuilt exchangePage Replacement Parts and Accessories ProbesTransfer Tubing and Waste Bottle Control SoftwareProbe Guide Inserts AppendixRack Accessories Rinse StationCables and I/O Accessories MiscellaneousRacks Code 23 rack Code 20 rackCode 21 rack Code 24 rackCode 30 rack Code 28 rackCode 29 rack Code 31 rackCode 32 rack Code 33 rackCode 200 rack Code 202 rack Code 201 rackCode 201H rack Code 203 rackCode 204 rack Code 205 rackCode 205H rack Code 208 rack Code 206 rackCode 207 rack Code 209 rackCode 212 rack Code 210 rackCode 211 rack Code 213 rackCode 217 rack Code 214 rackCode 216 rack Code 218 rackCode 505 rack Code 222 rackCode 228 rack Code 505H rackAppendix Peltier Racks Code 242 Peltier rackCode 542 Peltier rack Code 852 Peltier rack Code 853 Peltier rackCode 854 Peltier rack Gsioc Commands Gsioc Commands Gsioc Command ListCommand Type Description @aa=xxxx Bf,d Cv=ss..ss Exyz JccccpJcttt Ksxxx Oabcd Smm Ta,b,c,d Vcs..s Wcc..ccXax/y Xbx/y Xcx/y Xdx/y Yyyyy Zsp ZspB Pipetting Techniques Minimizing Cross-Contamination Minimizing Dilution Effect Appendix