Page 34

Auto

Use this function to perform an automatic sequence of tests on an inactive or active pair.

Press the [Up] or [Down] key to move to the Inactive or Active test. Press the [Enter] key to accept the choice go and to the measurement.

Inactive Pair

Inactive tests are performed on non-working lines and include: DC Volts, Resistance, Opens, Load Coils, and Capacitive Imbalance.

The results appear as each test is completed. If a test cannot be done, “***” appears in the appropriate result box.

34

Image 34
Contents Dynatel TM 965DSP-B Subscriber Loop Analyzer Page Table of Contents Out of Box Test Leads ScreenKeypad Control Keys EditingFunction Keys Setup Welcome ScreenUnits ContrastVoltage Current Resistance Toolbox Self-Calibration Load Coils Ohms/Distance Caller ID Opens Edit Custom Calibrate to Cable Tone Page RFL Resistance Fault Locate Separate Pair This is the preferred hookupPage RFL Single Pair Page RFL Setup RFL Multi-section SetupPage Isdn Isdn Bit Error Test Page Loss NoiseLongitudinal Balance Wideband LossAuto Inactive PairAuto Setup Active PairDial Noise/Longitudinal Balance Dial LossOpens Cable Type Talk Set Talk Set Setup Level of Charge Care and MaintenanceCharging AC/DC AdaptorReplacement Items Battery HolderTest Leads Symptom Cause Solution Unit does not turn on Repair965DSP-B Specifications New Product Limited Warranty