Talk Set

This function allows you to use the 965DSP-B as a Talk Set on active pairs, or to send DTMF tones or pulses on inactive pairs.

Connect the Red and Black leads to the pair for Loop Start. Also connect the Green lead to ground for Ground Start.

This screen displays the last number dialed, the DC voltage on the line, and the signal format for dialing. You can use DTMF or pulse dialing, or select between loop or ground start.

Use the [Setup] key to change these settings or to select and edit a stored phone number.

The blue keys become number keys in Talk Set. Use the [Backup] key to exit Talk Set before going to another function.

Use the [Tab] key to switch between on-hook and off-hook. Use the [Enter] key to dial (send) the number in the display.

Use the [Right] arrow key to turn the microphone on and off.

The speaker is active during the Talk Set function. Use the [Up] and [Down] arrow keys to adjust the volume. The maximum volume is limited in the Talk Set function to prevent unwanted feedback.

37

Image 37
Contents Dynatel TM 965DSP-B Subscriber Loop Analyzer Page Table of Contents Out of Box Test Leads ScreenKeypad Editing Control KeysFunction Keys Welcome Screen SetupContrast UnitsVoltage Current Resistance Toolbox Self-Calibration Load Coils Ohms/Distance Caller ID Opens Edit Custom Calibrate to Cable Tone Page Separate Pair This is the preferred hookup RFL Resistance Fault LocatePage RFL Single Pair Page RFL Multi-section Setup RFL SetupPage Isdn Isdn Bit Error Test Page Noise LossWideband Loss Longitudinal BalanceInactive Pair AutoActive Pair Auto SetupDial Noise/Longitudinal Balance Dial LossOpens Cable Type Talk Set Talk Set Setup Charging Care and MaintenanceLevel of Charge AC/DC AdaptorReplacement Items Battery HolderTest Leads Repair Symptom Cause Solution Unit does not turn on965DSP-B Specifications New Product Limited Warranty