ResMed S9 VPAP Tx Changing settings via the Home menu, Humidity level, Climate Control, Mask Fit

Page 16

Changing settings via the Home menu

From the Home menu, you can adjust or check the following features:

Humidity level

Humidity levels can be adjusted at any time to find the setting that is most comfortable for the patient.

Climate Control

When the ClimateLine or ClimateLineMAX heated air tubing is connected and Climate Control is enabled, the air temperature can be adjusted to find the setting that is most comfortable for the patient.

When set to Auto, Climate Control prevents rainout by maintaining 80% relative humidity in the delivered air. If Climate Control is set to Manual, Humidity Level and Heated Tube temperature can be set independently.

Mask Fit

Mask Fit is designed to help fit the mask properly to the patient.

The Mask Fit feature delivers CPAP pressure for a three-minute period, prior to starting treatment. During this time, the mask can be adjusted to minimise leaks.

To use Mask Fit:

1.Fit the mask as described in the mask user guide.

2.Press for at least three seconds.

One of the MASK FIT screens is displayed (as shown on the left).

3.If necessary, adjust the mask, mask cushion and headgear until there is a secure and comfortable fit. After three minutes, the pressure reverts to the set pressure and treatment will begin. You can end Mask Fit at any time by pressing .

12

Image 16
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions Welcome S9 Vpap Tx indications for useS9 Vpap Tx contraindications S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Masks Supplemental oxygenAntibacterial filters HumidifierTherapy mode Description Therapy modesS9 Vpap Tx basics Key FunctionSetting up the S9 Vpap Tx Filling the water tub English Mask type Settings Mask and air tubing settingsAir tubing Specifications Settings About the menus Navigating the menusHome menu Changing settings via the Home menu Humidity levelClimate Control Mask FitViewing the treatment screens Treatment screen parameters Parameter Modes DescriptionSetup menu Setup menu accessSettings ConfigurationClinical setup menu parameters Parameter ModesRange Description Ü ü ü ü ü ü ü üParameter Modes Range Description Ü ü üLED Info menu ServiceParameter Description Temperature Managing Climate ControlMode Heated tubeDelivering therapy Adding supplemental oxygen To add supplemental oxygenCleaning and maintenance Filter EfficiencyAlarms menu Working with alarmsTesting the alarms Troubleshooting Problem/Possible cause SolutionNo display Insufficient air delivered from the deviceAlarm is activated and the LCD screen display disappears Water splashing on your face from the H5iLeaking water tub Air feels too warm/cold in the maskGeneral technical specifications Technical specificationsS9 Vpap Tx technical specifications Pneumatic flow path Value Range Display resolution Flow maximum at set pressuresDisplayed values MinSymbols ServicingLimited warranty ProductPage GLO Multi